30
Display 8 - 0 vol.% adjustment
Display 8 - 0 vol.% adjustment
Display 8 - 0 vol.% adjustment
Display 8 - 0 vol.% adjustment
Display 8 - 0 vol.% adjustment
Now set the zero point of the vol.% range with
fresh air
fresh air
fresh air
fresh air
fresh air
using the
SPE 1
SPE 1
SPE 1
SPE 1
SPE 1
test set.
•
wait until the display has stabilised
•
confirm the adjustment with the
on/off button (
OK
OK
OK
OK
OK
appears in the
LCD display)
•
press the cursor-up button to
move to the next display
Display 9 - 100 vol.% adjustment
Display 9 - 100 vol.% adjustment
Display 9 - 100 vol.% adjustment
Display 9 - 100 vol.% adjustment
Display 9 - 100 vol.% adjustment
Now release the
100 vol.% methane CH
100 vol.% methane CH
100 vol.% methane CH
100 vol.% methane CH
100 vol.% methane CH
4
4
4
4
4
or
100 vol.% propane C
100 vol.% propane C
100 vol.% propane C
100 vol.% propane C
100 vol.% propane C
3
3
3
3
3
H
H
H
H
H
8
8
8
8
8
test gas
via the
SPE 1
SPE 1
SPE 1
SPE 1
SPE 1
test set. This requires a steel test-gas bottle with pressure
reducer and connecting hose.
•
wait until the display has stabilised
•
confirm the adjustment with the
on/off button (
OK
OK
OK
OK
OK
appears in the
LCD display)
•
turn off the test-gas feed
•
press the cursor-up button to
move to the next display