SRH400 Operating Instructions

10

Phone: (800) 257-3872  |  Fax: (978) 264-0292  |  www.setra.com 

8.3 1-point RH and T adjustment

For best performance, the RH and the T adjustment point 

shall be selected at the middle of the RH and T measurement 

range of main interest. 

1. Set the jumper to RH calib or to T Calib. 

2. Place the probe into the corresponding calibrator and let 

it stabilize at the selected RH or T value for at least 30 

minutes.

3. For adjustment point in the lower half of the output scale, 

press and hold S2 for 3 seconds. The LED “Calib” will light 

and the symbol “CAL<” will appear on the SRH400 display.  

For adjustment point in the upper half of the output scale, 

press and hold S1 for 3 seconds. The LED “Calib” will light 

and the symbol “CAL>” will appear on the SRH400 display.

4. Push S1 (up) or S2 (down) to bring the measured value in 

line with the reference in steps of 0.1. The actual corrected 

value is indicated on the display and can be measured at 

the analogue output.

5. Press and hold S1 for 3 seconds to store adjusted value 

in the S07 probe and end the first (low) point adjustment. 

The LED “Calib” will be deactivated and the display will 

show “CAL”.

Alternative: Press and hold S2 for 3 seconds to exit the 

adjustment mode without storing the adjustment data. 

The LED “Calib” will deactivate and the symbol “CAL” 

disappears from the LC display.

8.4 Return to factory calibration

During normal measurement mode of SRH400 (NOT during 

adjustment mode!) press and hold together S1 and S2 for 5 

seconds. The return to factory calibration is indicated by one 

flash of the LED “Calib”.

8.5 Loop adjustment and calibration in the field

Loop calibration or adjustment in the field, as required by the 

FDA (Food and Drugs Administration) regulated industries, 

is easily possible for the SRH400 with two separate probes. 

Use extension cables to drop the probes into calibrators 

without dismounting or disconnecting the SRH400 basic 

unit. The illustration shows the RH probe placed into the 

Humor 20 high end portable humidity calibrator, and the T 

probe in a dry block calibrator.

8.6 Probe individual adjustment

The probes can be adjusted individually with the optional 

Product  Configuration  Adapter  and  the  optional  Product 

Configuration Software.

9.0 Functional and accuracy check of 

SRH400 basic unit

The  SRH400 basic unit can be checked with respect to 

functionality and accuracy by using the SRH400 Reference 

Probes Set, order number SRH400XXX (Fig. 4).

The set consists of two reference probes, which supply each 

a pair of fix humidity and temperature values as follows:

RH = 10 % , T = 45 °C (113 °F)   

 

 

RH = 90 % , T = 5 °C   (41 °F)

The set includes a test report for each reference probe. 

For SRH400 basic unit dedicated for only one combined RH 

& T probe, both RH and T check shall be performed with each 

reference probe connected at the single probe connection 

socket. 

For SRH400 basic unit ordered with for two separate RH and 

T probes, the RH check shall be done with each reference 

probe  connected  to  the  first  (left)  socket,  and  then  the  T 

check with the same probe connected to the second (right) 

socket. 

Humidity calibrator

Temperature

calibrator

 RH probe

 T probe

Figure 3.

Loop calibration

Summary of Contents for SRH400

Page 1: ...Setra Systems Inc 159 Swanson Road Boxborough MA 01719 800 257 3872 www setra com Operating Instructions Model SRH400...

Page 2: ...t is for information purposes only and is subject to change without notice Setra Systems assumes no responsibility for any errors or for consequential damages that may result from the use or misrepres...

Page 3: ...8 0 RH and T adjustment of probes 8 1 Definitions 8 2 S07 adjustment with SRH400 basic device 8 3 1 point RH and T adjustment 8 4 Return to factory calibration 8 5 Loop adjustment and calibration in...

Page 4: ...led in a manner that does not have a negative effect on its safe use The user is responsible for observing all applicable safety guidelines local and international with respect to safe installation an...

Page 5: ...as on the optional display One or two point adjustment for RH and T of the transmitter canbeeasilyperformedwithpushbuttonsontheelectronics boardoftheSRH400basicunit Alternatively theprobescanbe adjus...

Page 6: ...H400 version with remote probe please see below 4 4 probes connected to SRH400 with extension cables SRH400 with remote sensing probe The probes can be also connected to the SRH400 housing with extens...

Page 7: ...e electrically isolated from the GND Locate the SRH400 the probe cable and the output cables as far as possible from sources of electromagnetic disturbances For the analogue outputs use either shielde...

Page 8: ...on indicates adjustment mode 1 flash confirms reset to factory calibration 4 S1 Functions 1 point RH or T adjustment RH 50 RH T within the upper half of the T scale 2 point RH or T adjustment high ca...

Page 9: ...hall remain constant Allow min 30 min stabilization time for RH calibration or adjustment If needed clean the sensors and replace the filter cap by a new one 8 2 1 2 point RH and T adjustment 1 Set th...

Page 10: ...S2 for 5 seconds The return to factory calibration is indicated by one flash of the LED Calib 8 5 Loop adjustment and calibration in the field Loop calibration or adjustment in the field as required b...

Page 11: ...15 35 V DC 15 29 V AC power supply 20 35 V DC RL 500 11 35 V DC RL 50 SRH400 Analog volts For the SRH400 with current output an ammeter shall be connected as in Fig 6 power supply for 0 1 V 10 35 V D...

Page 12: ...tput Iout mA 16 Ref Tmax 4 mA For voltage output Uout V Abb Ref Tmax Where Abb voltage scaling 1 V 5 V 10 V Ref reference value of temperature 45 C 113 F resp 5 C 41 F Tmax max temperature scaling e g...

Page 13: ...ers HA010103 SRHSS Stainless steel sintered filter Filters HA010104 SRHPG Plastic grid filter Filters HA010105 SRHPF PTFE filter 12mm Filters HA010106 SRHMG Metal grid filter Filters HA010115 SRHHP H2...

Page 14: ...To assure prompt handling please make sure the RMA number is on the outside of the box and a copy of the service request from is included in the shipment If applicable include a copy of the PO for re...

Page 15: ...s notified in advance of and the product is returned to SETRA transportation prepaid Unless otherwise specified in a manual or warranty card or agreed to in writing and signed by a SETRA officer SETRA...

Page 16: ...Setra Systems Inc 159 Swanson Road Boxborough MA 01719 800 257 3872 www setra com SS SRH400 Rev B 09 2018...

Reviews: