26
ENVIRONMENTAL CUSTOMIZATION
Most high precision balances are used in stable environments, away from
drafts and vibration. You may customize the balance with regards to its
environmental conditions. The balance’s response rate can be set to fast,
slow or auto, depending on your application.
The balance is factory set to the auto mode which is recommended for most
applications and will provide good stability and response.
The fast mode will provide a faster response than the auto mode and is
suitable for applications such as filling to a target weight. In environments
where draft and vibration are present, the displayed weight may not be as
stable as required.
The slow mode will provide maximum stability especially in a noisy or drafty
environment; however, the response rate may be slower.
To permanently set the display response rate for your environment or par-
ticular application, perform the following steps:
STEP 1.
TARE
POWER
MODE
FUnC
0.00
Hold down the
TARE
key then press the
MODE
key. Release both keys, the
display will read "FUnC".
STEP 2.
MODE
TARE
POWER
CUS7On
FUnC
Press the
TARE
key repeatedly until the display reads “CUSTOn” for the
customization menu.
Note:
To escape anytime during this procedure, press the
TARE
key until
“ESCAPE” is displayed and press the
MODE
key.