21
STEP 7.
MODE
TARE
POWER
-0.0648
0.0000
By holding down the
MODE
key, the displayed reading will increase. Holding
down the
TARE
key will decrease the reading. If the conversion number is a
factor (“x” units per gram), enter a positive number by holding down the
MODE
key until the number is displayed. If the conversion number is a divisor
(grams per “x” unit), enter a negative number by holding down the
TARE
key
until the number is displayed.
Note:
For example, to measure in grains per gram, calculate the conver-
sion; 1 grain is equal to 0.0648 grams and enter the divisor -0.0648.
The same result can be obtained using 15.43 grains/gram in which
case you would enter the positive factor 15.43.
STEP 8.
TARE
POWER
MODE
-0.0648
-0.0648
Once the conversion number is displayed, select it by pressing both the
MODE
key and the
TARE
key
simultaneously
.
STEP 9.
TARE
POWER
MODE
0.00
-0.0648
The number will flash on the display. If you need to adjust this flashing
number, press the
TARE
key and return to step 7. If the flashing number is
correct, press the
MODE
key. The balance will return to the normal weighing
mode and display the weight in “x” units.
Note:
If setting a user unit other than a flow rate (per second reading) the
flow rate must be turned off following the instructions in the next
section.