SERVOPRO FID Online Total Hydrocarbons Analyser
SERVOPRO FID 01000001A rev:5 43
CAUTION
To comply with EMC requirements, a screened cable must be used to
connect the mA output
.
CAUTION
To comply with EMC requirements, the cable connected to the mA output
terminal shall not be longer than 30 m, nor shall it leave the building in
which the analyser is installed without suitable isolation.
Connect the wires in your cable to the screw terminals on the rear terminal
connector (Figure 1). Section 2.7 provides information on the rating and size of
cable.
Figure 38 mA output connections
5.3.3 Digital outputs
WARNING
The output terminals are separated from the analyser mains circuits
by reinforced insulation. The terminals must only be connected to
circuits that are themselves separated from mains voltage by at
least reinforced insulation.
CAUTION
To comply with EMC requirements, a screened cable must be used to
connect the digital outputs
.