Installation, Start Up and Shut Down DF-760E
31
See the section on “Power Up Defaults on page 109 for setting user selectable preferences
at the time of power up.
4.2.4 Oxygen
Start the gas flow through the Oxygen Sensor by going to the Oxygen Controls Menu as
shown on page 72 and select Gas Valves. On the Gas Valves screen Figure 61, highlight
“Restore Process Gas Flow” and hit Apply, and then hit Done.
See the section on Gas Pressure and Flow Settings on page 29 for important information on
setting proper pressures and flow rates.
Allow the oxygen sensor to purge for 10 minutes before proceeding.
On initial start-up from the factory, the display will show “Sensor Off” indicating that the
oxygen sensor polarizing voltage is automatically turned off and must be manually turn on.
See page 72. Also, see the section on page 109 for further information on changing this
function on future power ups. After turning the oxygen sensor polarization voltage on, it
should take less than 5 minutes for the analyzer to come on scale (< 20ppm). The
concentration of oxygen is automatically shown in parts per million (ppm) or parts per
billion (ppb) on the display.
See the section on “Power Up Defaults” on page 109 for setting user selectable preferences
at the time of power up.
After analyzer start-up, the oxygen sensor can be used to check for leaks in plumbing
system. See page 120 for additional information.
NOTE
If oxygen is included in the background gas scale factor
calculation, the Oxygen sensor is automatically isolated. In
addition, no Zero adjustments will be possible using the internal
purifier on either the Oxygen or Moisture cells.
NOTE
Over-pressurizing the oxygen sensor can result in permanent
damage to the sensor. Limit the backpressure to the analyzer to
±1 psig.
4.2.5 Backflow Prevention System
It is imperative that flow be maintained through the analyzer. For this reason a flow
detection system is installed in the bypass loop to monitor the flow rate. In the event of
loss of sufficient flow the analyzer is automatically isolated to prevent ambient air from
contaminating the plumbing system. For additional information see page 106.
Summary of Contents for NanoTrace DF-760E
Page 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Page 4: ......
Page 38: ......
Page 44: ......
Page 122: ......
Page 128: ......
Page 138: ......
Page 142: ......
Page 143: ...Service DF 760E 139...
Page 144: ...140 DF 760E...
Page 148: ......
Page 166: ......
Page 168: ......
Page 171: ...DF 760E 167...
Page 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Page 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...