User Interface DF-760E
109
7.7.6 Power Up Default
Figure 123: Power Up Default
The Power Up Default SubRoutine enables the user to determine the default states during
analyzer power up. The power up states are useful because they determine whether, for
instance, the sensors are protected from ambient air contamination or whether they are
configured the best way for rapid station to station transfer and measurement.
Use the
NEXT
key to move from field to field. Use the up and down arrows (
and
)
to move between the selections within the field. When done use the
NEXT
key to move to
the
Accept
field and hit the
Enter
key. Pressing the
ESC
at any time will make no
changes and will return to the main data display
Figure 124: Power Up Default Screen
NOTE: If the analyzer auto-reboots due to a system error 20006, the power-up defaults are
ignored and the analyzer returns to the mode of operation found immediately before the
error. This reboot action is identified in the system error log as an error 222. See page 60
for additional information on system error codes.
Summary of Contents for NanoTrace DF-760E
Page 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Page 4: ......
Page 38: ......
Page 44: ......
Page 122: ......
Page 128: ......
Page 138: ......
Page 142: ......
Page 143: ...Service DF 760E 139...
Page 144: ...140 DF 760E...
Page 148: ......
Page 166: ......
Page 168: ......
Page 171: ...DF 760E 167...
Page 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Page 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...