![Servomex DF-310E Operator'S Manual Download Page 84](http://html1.mh-extra.com/html/servomex/df-310e/df-310e_operators-manual_1247052084.webp)
80 DF-310E
User Interface
Contact the factory, for assistance with gases not listed above.
When
GSF
is selected, the display in Figure 51 will be shown.
Figure 51: Gas Scale Factor
Entries for additional gases can be accessed by using the
or
keys to scroll through the
list. The entries spread across more than two screens. Continued pressing of
will give access
to the additional choices, shown in Figure 52. By moving the asterisk to the appropriate line and
pressing
, the volume percentage of the sample gas can be adjusted.
After the volumetric percent of the selected gas is entered, continue to press
until the number
is no longer in reverse video. Repeat the process for other gases in the sample gas composition.
Note: An error message will appear if the sum of gases does not equal
100
%. If that occurs,
change one (or more) values and press
again.
For percent oxygen Analyzers, assume oxygen has the same diffusivity as nitrogen. Thus, add
the percentage of oxygen to the percentage of nitrogen when entering the percentage of nitrogen.
At the bottom of the list, the display will show Figure 52 below.
Figure 52: Gas Scale Factor Menu (Cont’d)
Note: Scrolling down the gas list from Figure 51 to Figure 52 will displace one line at a time.
Because these figures are presented from the top and from the bottom of the gas list, H2
(Hydrogen) and NH3 (Ammonia) appear to be missing.
Summary of Contents for DF-310E
Page 1: ...Oxygen Analyzer DF 310E OPERATOR MANUAL...
Page 10: ......
Page 26: ...22 DF 310E Options Figure 7 Plumbing Configuration Options...
Page 36: ......
Page 62: ......
Page 66: ...62 DF 310E User Interface 3 43 ppm...
Page 96: ......
Page 108: ......
Page 118: ......
Page 122: ......
Page 124: ......
Page 129: ......