background image

Blue

Line 

Instruments for Electrophoresis 

 
 

SERVA Electrophoresis GmbH - Carl-Benz-Str. 7 - D-69115 Heidelberg 

Phone +49-6221-138400, Fax +49-6221-1384010 

e-mail: [email protected] - http://www.serva.de

 

 

 

 

 

I

NSTRUCTION 

M

ANUAL

 

___________________________________________________________________ 

 

BlueSpin Cryo 

(Cat. No. BSD-CRYO) 

 

 

 

 

Summary of Contents for BlueLine BlueSpin Cryo

Page 1: ...ctrophoresis GmbH Carl Benz Str 7 D 69115 Heidelberg Phone 49 6221 138400 Fax 49 6221 1384010 e mail info serva de http www serva de INSTRUCTION MANUAL ________________________________________________...

Page 2: ...Setting Time 10 4 4 Setting Temperature and Fast Cool 10 4 4 1 Setting Temperature 10 4 4 2 Fast Cool 11 4 5 Acceleration Deceleration 11 4 6 Program Saving Recalling 12 4 6 1 Saving 12 4 6 2 Recalli...

Page 3: ...he instrument could possibly result in personal injury or physical damage on the instrument or its accessories ALWAYS locate the instrument on a flat rigid and stable table capable of withstanding the...

Page 4: ...trifuge rotor bucket or adaptors and sample container especially the guaranteed g force of sample container should not be neglected The rotors should be cleaned and kept dry after every use for longer...

Page 5: ...30 x 2 0 ml tubes 8 x 8 tube PCR strips Temp range C 20 40 FAST COOL button Yes Time control Pulse timed 100 min or continuous RPM RCF conversion Yes Noise level 56 dB Acc Dec 9 10 steps Program memor...

Page 6: ...rotor recognition function When you supply the power the display shows Searching Rotor Change to Rotor ID If the rotor is absent Error 9 will be displayed This message will be cleared after rotor cou...

Page 7: ...you don t close the rotor lid completely it will be crushed 10 Positioning of sample tubes Before loading sample tubes check for water drops or dirt in the rotor hole or inner adaptor If there is a w...

Page 8: ...7 Correct Ways of Sample Balancing Tube usage...

Page 9: ...acceleration deceleration level from 1 to 9 steps 0 in deceleration step means natural deceleration Larger number means faster acceleration or deceleration PROG Use to save a set of setting values or...

Page 10: ...uch the button for 5 seconds the setting mode is cleared Touch the Enter button to save and complete the setting 4 2 2 Setting RCF 1 Speed setting unit 1 rcf or 10 rcf RCF MODE is generated by pressin...

Page 11: ...o pass the sec value setting After keeping holding finger on the buttons for 5 seconds the unit of setting value is changed from 1 sec to 10 sec If you do not touch the button for 5 seconds the settin...

Page 12: ...e is safe during spinning at 1 000 rpm Before starting Fast Cooling please check the rotor coupling and symmetry of sample tubes 4 5 Acceleration Deceleration 1 Touch ACC DEC button 2 Touch the button...

Page 13: ...ram just touch the PROG button shortly less than 1 sec 2 Touch the buttons to select program number you want to recall and then touch the Enter button If you do not touch the button for 5 second the s...

Page 14: ...or is released Manual opening should be performed only when spinning is completely stopped Otherwise harmful damage will be accompanied to not only operators but samples After opening the door manuall...

Page 15: ...he motor shaft to avoid any imbalance problem due to the contaminants After using the instrument take out the rotor from the shaft and clean the shaft with dry soft cloth to keep dry 5 4 Rotor If any...

Page 16: ...ortly open the door with spikes for safety of sample Can t close the door Remove the dirt at the door latch and then close the door completely again If the door seems not being closed by mechanical re...

Page 17: ...or is overheated Shut off the power supply for an hour and then turn on the power switch for checking the instrument If the error code shows continuously please contact your distributor Error 4 Low Vo...

Page 18: ...s not reached to setting temperature within an hour No user action Please contact your distributor Error 12 Temp Sensor Error There is a faulty in the temperature sensing of chamber or over heated No...

Page 19: ...with aluminum lid for 24 microtubes BS MRR24 A Fixed Angle Microtube Rotor with aluminum lid for 30 microtubes BS MRR30 A Fixed Angle PCR tube Rotor for 8 rows of PCR strips no lid BS MRP64 0 2 ml Ada...

Page 20: ...nts of the EC EMC Directive 2004 108 EC 21st May 2007 EC Low Voltage Directive 2006 95 EC 12th December 2006 is This declaration is no longer valid if the machine is modified without SERVA Electrophor...

Page 21: ...1 13840 44 Fax 06221 13840 54 E Mail tech service serva de Technical Service Collagenase Phone 04122 712 413 Fax 04122 712 286 Free Phone 0800 737 8246 Free Fax 0800 737 8247 International Customers T...

Reviews: