
SeQual Technologies Inc.
Part No: 5167
7
Alarm System
- The oxygen concentrator has an alarm system with a red visual alarm indicator light and
buzzer. The alarm system will activate in the event of an internal fault or power outage. The
red and/or yellow light may stay on for up to ten (10) minutes after start-up, OM models only.
This is a normal condition during the initial warm-up period.
If the oxygen concentrator is equipped with an oxygen concentration monitor (OM model), this monitor
indicates if the oxygen concentrator is performing to the correct specifications. At normal oxygen
concentration levels, the green light is activated. At lower than normal oxygen levels, the yellow light is
activated. The red light activates and the unit alarms if the system malfunctions or when service is needed.
The oxygen concentrator will continue to run at lower levels of oxygen but may shut down under other
system malfunctions. If the red light turns on, you should discontinue use of the oxygen concentrator and
refer to the troubleshooting section of this manual.
Operating Your Oxygen Concentrator
This manual serves as a supplemental reference to help you operate and maintain the unit. If you have any
further questions or problems call your SeQual dealer or authorized service representative.
General Precautions
– Retain the product’s safety and operating instructions for future reference. Follow all
operating and usage instructions. Observe all warnings on the product and in the operating instructions. To
reduce the risk of fire, bodily injuries, and damage to the oxygen concentrator, observe all safety precautions.
Do Not
connect to an extension cord or electrical outlet controlled by a switch. No other
appliance should be plugged into the wall outlet. Voltage requirements are marked on the label
on back of unit.
Important:
Read “Important Safety Guidelines” section of this manual before plugging-in the power cord.
Do Not
use the unit if the power cord is damaged
.
Warning:
Improper use of the power cord and plug may result in a fire or electrical shock hazard.
Check to see if the power cord is plugged in and there is power to the outlet.
Check that the oxygen concentrator is not placed near a heat register.
Important:
SeQual’s products operate safely when used according to their marked electrical ratings and product
usage instructions. The power cord used with the Regalia Oxygen Concentrator must meet the requirements of
the country where the product is used.