Recommended Operating Environment Guidelines
Temperature
50
º
F to 104
º
F (10
º
C to 40
º
C)
Electrical
Use no extension cords or electrical outlets controlled by a switch.
Placement
No closer than 3 inches (7.6 cm) from the wall, draperies and furniture.
Environment
Smoke, pollutant, and fume free. Relative Humidity: 10% - 95%
Operating Time
Up to 24 hours per day.
Flow Rate
From 0.5 LPM to 10.0 LPM (for the 10 LPM version)
From 0.5 LPM to 7.0 LPM (for the 7 LPM version)
0
0
30
35
40
55
60
65
70
75
80
85
90
95
Relative Humidity, %
Te
m
p
e
ra
tu
re
,
°C
Important:
For rated performance at 10 LPM (or 7 LPM for the 7 LPM version) do not operate the Oxygen
concentrator above the indicated boundaries. Operating at flow rates less than 10 (or 7 LPM for the 7 LPM
version) LPM will tolerate higher temperature/humidity conditions.
The Proper Location
Usage in environments other than those described on the previous page may result in the need for increased
oxygen concentrator maintenance. Select a location in a well ventilated space that avoids fumes and pollutants
and allows the oxygen concentrator to draw in room air through the intake filter on the top of the cabinet without
being restricted. Make sure the oxygen concentrator is at least three to six (3 - 6) inches (7.6 – 15.2 cm) away
from walls, furniture and especially curtains or draperies that can impede or block adequate airflow to the oxygen
concentrator.
Do Not
cover unit with a blanket, towel, quilt, etc.
Do Not
place objects under unit. Maintain at
least ½ inch clearance under unit. Keep the oxygen concentrator at least five (5) feet (1.5 m) away from hot,
sparking objects or open sources of flame.
Do Not
locate the unit near any heat source or heat register.
Before Operating Your Oxygen Concentrator
There are many different types of oxygen tubing that can be used with this oxygen concentrator.
Oxygen Tubing - C
onnect the oxygen tubing directly to the oxygen outlet port. Connect the other end of the
tube to the application, if not already attached.
Important:
The tube fitting may be tight
-
Do Not
use oil or
grease to lubricate
.
To ensure there is enough pressure to deliver oxygen, the length of the tubing should not
exceed 50 feet (15.2m). Maximum extension tubing length is dependent upon environmental
conditions, tubing type, liter flow and any accessories used between DISS fitting and
application. Shorter tubing lengths will be required in some cases. The internal diameter of
the tubing should be no less than 3/16-inch (0.48 cm).