SeQual Regalia Instruction Manual Download Page 10

 

 

SeQual Technologies Inc.    

  

Part No: 5167 

10 

2.

   Display of Operating Conditions 

Flow Rate 

Shown? 

Green 

Light 

Yellow 

Light 

Red 

Light 

Audible 

Alarm 

Unit 

State 

Operating Condition 

No 

Off 

Off 

Off 

Off 

Not 

Energized 

Unit not running / Flow 

Indicator Display Not Lit 

No 

Off 

Off 

Off 

Off 

Energized 

Unit not running / Flow 

Indicator Display Lit 

Yes 

On 

On 

On 

On for 3 

seconds 

Started 

Running 

Unit is Starting 

Yes

 

On 

On 

On 

Periodic 

Beep 

Running 

Unit is warming up* 

Yes

 

On 

On 

Off 

Periodic 

Beep 

Running 

Yes

 

On 

Off 

Off 

Off 

Running 

Normal Operating 

Condition  

Yes

 

Off 

On 

Off 

Periodic 

Beep 

Not 

Running 

Unit turned 

On/Off/On Quickly 

Yes

 

Off 

Flashing 

Off 

Off 

Running 

Oxygen Concentration 

Below Normal Level * 

Yes

 

Off 

Off 

Flashing  

On  

Running 

Oxygen Concentration 

Abnormal Level *

1

 

Yes

 

Off 

On 

Off 

Off 

Running 

Button Stuck 

Yes

 

On 

On 

Off 

Off 

Running 

9V Battery Low Voltage 

(if enabled) 

No 

Off 

Off 

Flashing 

On 

Not 

Running 

Unit is overheating 

Compressor Malfunction 

Power Interruption 

Internal Fuse Blown 

No 

On or 

Off 

On or 

Off 

On 

On 

Running 

System Malfunction

2

 

* These conditions are displayed only on units equipped with oxygen monitor. 

1

 If abnormal conditions are detected and cannot be remedied, refer to Troubleshoot 

section of this manual.

 

2

 During System Malfunction condition the Green and Yellow lights may be On or Off.

 

 

 

3.

 

Non-Usage - 

When not using the oxygen concentrator, press the On/Off button to turn it off.  

 

Summary of Contents for Regalia

Page 1: ...Regalia Oxygen Concentrator REGALIA INSTRUCTION MANUAL P N 5167 Rev A March 2006 ...

Page 2: ...nment Guidelines 6 The Proper Location 6 Before Operating Your Oxygen Concentrator 6 Operating Your Oxygen Concentrator 7 Installation Instructions 8 Select a Location 8 Inspect Inlet Air Filter 8 On Off Button 8 Warm up Period 8 Flow Rate Adjustment 8 Power Management Feature 9 Operating Conditions 9 Control Panel Display 9 Display of Operating Conditions 10 Non Usage 10 Maintenance 11 Trouble Sh...

Page 3: ...void the warranty Use No Oil or Grease A spontaneous and violent ignition may occur if oil grease or other petroleum substances come into contact with oxygen under pressure Keep these substances away from the oxygen concentrator tubing and connections and any other oxygen source Do Not use any petroleum based or other lubricants unless recommended by SeQual Technologies Inc Keep oxygen concentrato...

Page 4: ...ng information specific instructions commentary sidelights or interesting points Symbols Used in Instruction Manual and on Oxygen Concentrator Refer to this section for an explanation of the symbols and the warnings that accompany them Read Instructions This symbol when used alone or in conjunction with any of the following symbols indicates the need to consult the operating instructions provided ...

Page 5: ...of protection against electrical shock than provided by type B applied part Class II Equipment Equipment in which protection against electrical shock does not rely on basic insulation only but in which additional safety precautions are provided Amperes Alternating Current Oxygen Concentrator Information SeQual s Regalia Oxygen Concentrators are quiet reliable and low maintenance sources of oxygen ...

Page 6: ...ake filter on the top of the cabinet without being restricted Make sure the oxygen concentrator is at least three to six 3 6 inches 7 6 15 2 cm away from walls furniture and especially curtains or draperies that can impede or block adequate airflow to the oxygen concentrator Do Not cover unit with a blanket towel quilt etc Do Not place objects under unit Maintain at least inch clearance under unit...

Page 7: ...ference to help you operate and maintain the unit If you have any further questions or problems call your SeQual dealer or authorized service representative General Precautions Retain the product s safety and operating instructions for future reference Follow all operating and usage instructions Observe all warnings on the product and in the operating instructions To reduce the risk of fire bodily...

Page 8: ... will illuminate to indicate the unit is running 4 Warm up Period After initially turning on the unit allow three 3 to ten 10 minutes for the unit to reach its specified performance If the unit is equipped with an oxygen concentration monitor OM models only the red yellow and green light will intermittently illuminate while the unit is warming up For details on warm up refer to the Display of Oper...

Page 9: ...tion Important During the power interruption The user can press the ON OFF button to silence the alarms and turn the unit off If the unit was not turned off the Regalia will automatically resume operation once power is restored If the unit was turned off during the power interruption the user must press the ON OFF button again after the power is restored to turn the unit on Operating Conditions 1 ...

Page 10: ...n Off On Quickly Yes Off Flashing Off Off Running Oxygen Concentration Below Normal Level Yes Off Off Flashing On Running Oxygen Concentration Abnormal Level 1 Yes Off On Off Off Running Button Stuck Yes On On Off Off Running 9V Battery Low Voltage if enabled No Off Off Flashing On Not Running Unit is overheating Compressor Malfunction Power Interruption Internal Fuse Blown No On or Off On or Off ...

Page 11: ...e air inlet filter is the most important activity you will perform to keep your oxygen concentrator performing properly Check this air filter daily and clean it at least once per week 1 Remove the filter from the cabinet 2 Wash the filter in warm water using a mild detergent solution 3 Rinse the filter thoroughly and squeeze out the excess water 4 Allow the filter to air dry Caution The filter sho...

Page 12: ...ack Light On Unit overheating Remove any obstruction s from air openings and refer to page 6 The Proper Location Compressor malfunction Service required Table 2 Concentrator Running Symptom Probable Cause Remedy No oxygen flow see Alarm Indications in Table on Page 11 Tubing blocked or kinked Clean or replace tubing Remove tube kink Air Inlet Filter blocked Clean air inlet filter Internal componen...

Page 13: ...ion Excessive Internal Temperature Outlet Pressure 35 36 psig 241 2 248 1 kPa relief valve Low Oxygen Output 70 oxygen Back up Alarm Power 9V Battery Filters Cabinet Air Inlet Compressor Inlet and HEPA Device Classification IEC Class II Type BF Applied Part Continuous Operation Transport Storage Requirements Temperature 4 F 20 C to 140 F 60 C Humidity Up to 95 Non Condensing Back pressure Effect o...

Page 14: ...ul interference to other devices which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving device Increase the separation distance between the equipment Connect the equipment into an outlet on a circuit different from that which the other device s are connected Co...