26
5.6.3 NIST buffer set
The table containing the temperature dependence of the NIST buffer set as it is used in the SI meter is
presented in Table 3 below.
°C
°F
#1
#2
#3
#4
#5
5
41
1.67
4.00
6.95
9.40
13.21
15
59
1.67
4.00
6.90
9.28
12.81
25
77
1.68
4.01
6.87
9.18
12.45
35
95
1.69
4.02
6.84
9.10
12.13
45
113
1.70
4.05
6.83
9.04
11.84
55
131
1.72
4.08
6.83
8.99
11.57
65
149
1.73
4.11
6.84
8.94
11.33
75
167
1.75
4.15
6.85
8.90
11.11
85
185
1.78
4.18
6.87
8.88
10.91
95
203
1.81
4.23
6.89
8.83
10.73
Table 3: NIST Buffer set
5.6.4 JIS buffer set
The table containing the temperature dependence of the JIS buffer set as it is used in the SI meter is
presented in Table 4 below.
°C
°F
#1
#2
#3
#4
#5
5
41
1.67
4.01
6.95
9.39
13.21
15
59
1.67
4.00
6.90
9.27
12.81
25
77
1.68
4.01
6.86
9.18
12.45
35
95
1.69
4.02
6.84
9.10
12.14
45
113
1.70
4.04
6.83
9.04
11.84
55
131
1.72
4.08
6.84
8.99
11.58
65
149
1.73
4.11
6.84
8.94
11.33
75
167
1.75
4.14
6.85
8.91
11.11
85
185
1.78
4.18
6.87
8.87
10.91
95
203
1.81
4.23
6.89
8.83
10.73
Table 4: JIS Buffer set