SI line pH PROBE MANUAL 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

For products: 
ConeFET 

3280-010 

CupFET 

3200-010 

LanceFET 

2270-010 

LH  2274-010 
MicroFET 

9270-010 

MiniFET 

9202-010 

Summary of Contents for ConeFET

Page 1: ...SI line pH PROBE MANUAL For products ConeFET 3280 010 CupFET 3200 010 LanceFET 2270 010 LanceFET H 2274 010 MicroFET 9270 010 MiniFET 9202 010...

Page 2: ...he probe as this may cause damage and will void the guarantee All information provided in this manual is applicable at the moment of publication Sentron holds the right to change components functions...

Page 3: ...of compliance 4 1 2 Guarantee 4 1 3 Specifications 5 2 Use 6 2 1 Installation 6 2 2 Preparing for use 7 2 3 Operating tips 8 3 Cleaning 9 3 1 Daily cleaning 9 3 2 Revitalizing 9 3 3 Cleaning tips 10...

Page 4: ...n will according to its own judgment either replace or repair a defect probe within 6 months of shipment provided the cause is a defect in material of manufacturing and the defect appeared during corr...

Page 5: ...5 1 3 Specifications...

Page 6: ...obe failure or irreversible damage to the probe Samples must be aqueous solutions or semi solids and compatible with the probe s materials If information is required regarding the chemical resistance...

Page 7: ...page from a new probe is normal and will not affect the lifetime or performance of the probe Subsequently rinse with deionized or distilled water Soak the probe in pH 7 00 buffer or pH 4 00 buffer wit...

Page 8: ...late based are less susceptible to carbon dioxide contamination than pH 10 00 buffers borax or carbonate based When slope errors occur it usually indicates a failing probe or a contaminated buffer If...

Page 9: ...e to be cleaned with a brush Place the probe in warm tap water around 60 C 140 F with a mild detergent for 5 minutes stir periodically Scrub the probe tip with a soft brush and water with a mild deter...

Page 10: ...r Cleaning agents are available at Sentron ask your local dealer Do not use hydrofluoric acid acetone MEK or similar agents to clean the probe 4 Storage For short time periods 2 Days probes can be sto...