System Setup
Probe Configuration
Copyright © 2008-2012 Sentek Pty Ltd All rights reserved
Page 8
Important:
VIC water counts must be taken in reverse osmosis (RO) water (EC less than
300
μ
S/cm). Always normalise sensors under the same conditions to ensure maximum
repeatability.
4. If required, enter custom calibration coefficients for each sensor. Each coefficient should be separated by a
semicolon (i.e. A;B;C). See the Sentek Calibration Manual for more information.
5. Save these settings to the probe by clicking
Write to Probe
.
Figure 5: Configuration tab
Clock tab
1. Click on the
Clock
tab to display the clock settings. Decide how often the probe should take a reading and set
the desired sampling interval accordingly.
WARNING:
Probes can only store about 2000 samples, after which the oldest samples are discarded. You
must download before this capacity is reached, otherwise new samples will overwrite the oldest
samples. The number of days before overwriting is dependent on the sample interval. A
scenario of 10 minute sampling gives approximately 14 days before a download is required.
2. Set the probe clock to the desired time (usually probe local time) by either typing it in or synchronizing it with
your computer clock.
3. Write to Probe