MA-DISP-FP4x & 6x_EN
Page 15 on 18
Rev: 08/02/2012
11.
EQUIPMENT SPECIFICATIONS
Case Material
Heavy duty welded uPVC
Connectors
Internal detachable Screw Terminal connectors accessed via compression
glands
Environmental
Storage Temperature range: -20 to +70C, non condensing
Operating temperature range: 0 to 50°C.
Power
100-240VAC, 45 to 60Hz or 11-30VDC optional
Burden
40VA maximum
Sealing
IP65 all round, provided the display is mounted vertically and that all cable
glands and rear case-closure screws are properly secured.
Input Signals (bipolar)
0-10V, 0-5V, 1-5V, 0-10mA, 0-20mA, 4-20mA
Input Resistance
1 MΩ for voltage, 33Ω for current inputs
Accuracy
+/-0.05% of range
Span temperatures drift
25 ppm/Degree Celsius
Zero temperatures drift
30 ppm/Degree Celsius
Filtering / smoothing
Selectable time constants of 0 to 25 seconds.
A/D conversion
Sigma-Delta 10 conversions per second, 50/60Hz rejection
Resolution
1 in 400 000 max. over full range
Display update rate
10 readings per second.
Display Range (max)
FP4x
-1999 to +9999, depending on available signal level.
FP4x
-199999 to +999999, depending on available signal level.