METIS M308 / M309 / M313 / M316 / M318 / M323 / H309 / H316 / H318 (12-pin)
SensorTools Software
32
Sighting settings
(device specific)
:
-
the
laser targeting light
can be switched on and off.
CAUTION, Laser Radiation, Laser Class 2:
Never look into the direct or reflected laser beam.
Do not point the laser to anyone.
If laser radiation hits the eye, the eyes must be intentionally closed and the head
immediately moved out of the beam.
-
with
camera module
monitor settings are configured (see
Control window
-
With a
motorized-focus optics
, the measurement or focus distance can be adjusted (see
7.2
Software Windows
7.2.1 Main Window
Graphs
The graph window in the main window shows the temperature profile in time.
Scale X-Axis
allows to set the time at which the graphs
window is filled with measured values.
If the time is changed in "Scale X-Axis", the graph is rebuild from the center of the window.
rebuilds the graph from the left.
/
Switches the X axis of the graph between time and a stopwatch where with pressing
the displayed time can be set to 0 at the current graph position.
Opens a
control window
for each connected device, if they have not been activated in the
connecting window (devices overview tab).
Select display graphs (right click
to change the color settings of
the graph), activates or deac-
tivates graphs
Time axis settings
Measured temperature
numerical / graphical
Measurement settings
Temperature scale
Main window
Control window
Information to sample time,
control time and slope time
Record
Extended measure-
ment settings
status of the digital
inputs / outputs
Time axis (current
time) / manual 0
Homepage link