METIS M308 / M309 / M313 / M316 / M318 / M323 / H309 / H316 / H318 (12-pin)
Mechanical Installation
20
5.6.2 Manually Adjustable or Motorized Focus Optics
with camera module
Optics
Model
Temperature ranges
[°C]
M308:
all tempera-
ture ranges
M318:
100-700°C
H316:
ZSC <400°C
H318:
120
–520°C
M309:
all ranges
M316:
all ranges
M318:
150
–1200°C
180
–1300°C
H309:
all ranges
H316:
ZSC
≥400°C
H318:
180
–800°C
Meas.
distance
a [mm]
Optics
pull-out
S [mm]
Spot size diameter M [mm]
M308 / M309:
OV09-D1
M316 / M318:
OV09-D2
340
26
2.3
1.8
0.9
400
20,9
2.5
2.2
1
500
15,6
3
2.5
1.2
600
12,5
3.8
3
1.5
800
8,6
5.5
4.4
2.2
1000
6,5
7
5.6
2.8
1500
3,8
8.8
7
3.5
2000
2,5
11.8
10
4.7
2500
1,8
13
10.4
5.2
3000
1,3
16.5
14
7.5
3500
0,9
18.5
15.5
8
4000
0
22
19
11
ZSC = Zero-scale temperature
Aperture diameter D:
Full scale temperature value
≤ 1400 °C: 14–16 mm;
> 1400 °C: 7
–8 mm
(D is dependent on the optics pull-out: optics pulled out: higher value, optics inserted: smaller value)
5.6.3 Manually Adjustable or Motorized Focus Optics Model M313
Optics
Model
Temperature ranges
[°C]
M313:
500
–2600°C
550
–3000°C
600
–3500°C
M313:
450
–1800°C
500
–2200°C
M313:
400-1400°C
Meas.
distance
a [mm]
Optics
pull-out
S [mm]
Spot size diameter M [mm]
OM09-C0
340
26
0.7
0.8
1.5
400
20.9
0.8
1.1
1.8
450
17.9
0.9
1.3
2.1
500
15.6
1.2
1.4
2.4
600
12.5
1.4
1.7
2.9
700
10.5
1.6
2.1
3.6
800
8.6
1.9
2.5
4.1
900
7.6
2.1
2.6
4.6
1000
6.5
2.4
2.9
5.3
1500
3.8
3.7
4.6
7.9
2000
2.5
4.9
6.3
10.7
2500
1.8
6.3
7.9
13.9
3000
1.3
7.6
9.5
16.8
3500
0.9
8.8
11.0
19.4
4000
0
10.5
12.6
23.1
Aperture diameter D:
Full scale temperature value
≤ 2200 °C: 14–16 mm; > 2200 °C: 7–8 mm
(D is dependent on the optics pull-out: optics pulled out: higher value, optics inserted: smaller value)