7-Examples & Interpretation
33
7
Examples & Interpretation
Due to space considerations, only the depth slices with discernible features are displayed for
each case study.
Case #1
This grid scan was conducted on the 5
th
floor of a medical building that was undergoing some
renovations. The grid size was 600x600mm, normal resolution.
In Figure 7-1a, we see rebar located between 60-90mm depth. If we go deeper to 240-270mm
(Figure 7-1b), there are two conduits running obliquely across the grid area. In the slice
immediately below that (Figure 7-1c), there are some features at numeric lines 4 & 6.
a)
b)
c)
Figure 7-1: