4.4 Display:
1. When the function is activated, the display shows and twinkles the description, “Clean Running”.
At the same time, the ACT indicator LED lights up, and the transmitter automatically turns off Relay 1
and Relay 2 function. After finishing cleaning, the Relay 1 and Relay 2 will automatically back to
normal status.
2. When Relay 1/Relay 2 which is set in high setting point is in action, the display shows and flashes the
description, “REL 1-HI/REL 2-HI”, and ACT indicator LED lights up. When Relay 1/Relay 2 which is
set in low setting point is in action, the display shows and flashees the description, “REL 1-Lo/ REL
2-Lo”, and ACT indicator LED lights up.
3.When the Analog 1 current output exceeds the upper/lower limitation, the display flashes ”pH-mA /
pH-mA ” or ”ORP-mA / ORP-mA “.
4. When the Analog 2 current output exceeds the upper/lower limitation, the display flashees “
˚
C-mA /
˚
C –mA ”.
Analog 2 output
current over range
alarm
Control function
on hold
:
Measurement mode
:
Set-up mode
:
Calibration mode
REL
1
high
or
low point alarm
Measurement unit
Clock
REL 2 high or
low point alarm
Analog 1 output
current over range
alarm
Temperature
compensation mode
(MTC/ATC)
Temperature value
Measurement value
▲
▲
▲
▼
▼
▼
Summary of Contents for TX2000
Page 1: ...1 T X 2 0 0 0 Intelligent pH ORP Transmitter Operation Manual KϵϬϬϭͲ2008CertificateNo C0036132 ...
Page 9: ......
Page 10: ......
Page 25: ...3 ...
Page 30: ...4 ...