3.3.2

 

Automatic accumulation 

Automatic accumulations do not require the INTRO key to be pressed for weighing; the unit 
simply  has  to  detect  a  stable  weight  value  greater  than  the  value  defined  as  minimum 
weighing (*) after the last weighing has been discharged. 

 

 

 

 

3.3.3

 

Totalise 

To  complete  and  totalise  all  weighings  in  progress,  keep  the  FUN  and  ENT  keys  pressed 
down  at  the  same  time.  When  doing  this,  the  screen  display  switches  between  the  total 
number of weighings made and the total weight accumulated, returning to weight mode by 
pressing any key. 
 
For example, if 4 weighings are carried out with a total weight of 0.632 kg, the unit switches 
between the number of weighings 
 

 

and the total accumulated weight 
 

 

until any key is pressed and the unit returns to weight mode. 
 

Automatic weighing mode is selected using the AUTOAC option in the 

Functions menu. 

 

(*) The minimum weight value for a weighing is determined using the 

WEI.MIN option in the unit's WE.CTR. programming submenu. 

 

Summary of Contents for SC-A1

Page 1: ...SC A1 Manual User SC A1 Advanced weighing equipment with ABS or INOX finish high contrast LED or backlit LCD display and 5 key keyboard 2017 02 14 COMPLETE USER MANUAL...

Page 2: ...2...

Page 3: ...on of serial weighings Automatic totalling of serial weighings Accumulated Grand Total Configurable number of lines to finish ticket and delays per printed line Repeat ticket and optional ticket per w...

Page 4: ...11 3 1 2 Stability 11 3 1 3 Quick tare 12 3 1 4 Gross weight net weight 13 3 1 5 Tare lock 13 3 2 Manual tare 14 3 3 Extended accumulation and totalisation of weighings 15 3 3 1 Manual accumulation 15...

Page 5: ...5 1 DISPLAY AND KEYBOARD DESCRIPTION Stability Zero Tare Tare lock Gross Piece counter Non metrological Low battery ON OFF On Off T Tare B Zero Gross FUN Function INTRO Enter Print...

Page 6: ...l is completed FUN Leave INTRO Save and exit Entering data digit by digit T Increase digit B Change the digit to be modified indicated by the corresponding decimal point FUN Leave INTRO Save and exit...

Page 7: ...o NU ACUM Number of accumulations before automatic totalling 0 255 accumulations W UNIQ Single weighing mode selection Yes No There will be a complete ticket for each weighing and also when totalising...

Page 8: ...INTRO to save FUN to leave DATE Displays the current date DD MM YY INTRO to edit T B to modify 6 decimal digits INTRO to save FUN to leave CELLS Display points and weight of each load cell Only in equ...

Page 9: ...s digits and pilots The equipment identifier is then displayed Then if the equipment has dead weight units with weight module or digital cell communication the counter value is shown the number of tim...

Page 10: ...Ordinary weighing Net gross tare tare lock By defect the unit will start up showing the weight on the platform Both the text of the banner displayed when starting up the unit and the headers and foot...

Page 11: ...tions The speed with which the changes and the degree of weight stability shown are displayed depend on the values set in the filter parameters Options STAB T FILTER and FIL DEP in the Programming Men...

Page 12: ...zero pressing the T key again causes the unit to carry out a new Tare The Quick Tare function it s only available when the number of divisions of gross weight is bigger than the one defined in the min...

Page 13: ...ght is removed from the platform deleting and re displaying the gross weight 3 1 5 Tare lock To prevent tare from being disabled when all the weight is removed from the platform press the T key a seco...

Page 14: ...d leave without validating and the ENTER key to validate the tare value entered taring the unit whilst locking the Tare When a manual tare is active printing the tare on a ticket will be accompanied b...

Page 15: ...y out weighing there must be stable weight which exceeds the value defined as minimum weighing with the weight dropping below the minimum weight necessary to carry out weighing at some point since las...

Page 16: ...NT key to display the message The unit then briefly displays the number of weighings accumulated so far in our case the first And the accumulated weight subtotal which in our case since it is the firs...

Page 17: ...s a frame to be sent with the weight on the platform at this moment In order to carry out a new weighing and continue accumulating remove the weight from the platform and place the new weight to accum...

Page 18: ...en doing this the screen display switches between the total number of weighings made and the total weight accumulated returning to weight mode by pressing any key For example if 4 weighings are carrie...

Page 19: ...t of 0 081 kg has been entered and accumulated Totalised by pressing FUN ENT 3 4 Piece counter In piece counter mode the unit calculates and displays at all times the number of pieces resulting from d...

Page 20: ...it can switch between weight and piece counter modes by pressing the FUN key The unit enables the pilot to indicate that the value shown on screen corresponds to a number of pieces and not to a weight...

Page 21: ...se the value of the flashing digit the B key to change digit the FUN key to cancel editing and leave without saving and the INTRO key to validate and save the piece value entered The unit will very br...

Page 22: ...NT The ticket resulting from the operations with pieces has the following format This indicates the unit weight of the pieces the weight and number of the pieces of each weighing and the total totalis...

Page 23: ...weighings so that the information relative to the latest 500 000 weighings can always be recovered The information of the memory is organized depending on the number of Ticket to which the weighing be...

Page 24: ...played Once the ticket number has been introduced the equipment searches on the DSD memory the relative data to this ticket In case it does not find any associated weighing the following message is di...

Page 25: ...5 and the total accumulated weight By pressing T and B keys we will be able to move through the various weighings First of all the number of weighing is displayed and afterwards the recorded net weigh...

Page 26: ...by pressing INTRO key the printing of the ticket copy is initiated at the same time that the flashing message Print is displayed to indicate that a printing is being made Finally FUN key exits the dat...

Page 27: ...27 Electromagnetic susceptibility has been certified in tests conducted with field strength of 10 V m in accordance with new regulations...

Reviews: