39
USER MENU
SHOw BH LOG
From the working display access the menu by pressing and holding the
TICK/MENU BUTTON (B) until the display reads USER MENU SHOW
TIME. Press the SAVE/ZERO BUTTON (C) to scroll to USER MENU
SHOW BH LOG.
Press TICK/MENU BUTTON (B) to select this feature. The last record
will be displayed. The heading will display BH LOG: XXX (indicating
record number) and the date/time of the record. Below will include the
recorded (PK) and (ON) concentrations. The SAVE/ZERO BUTTON (C)
advances to the next record while pressing the TICK/MENU BUTTON
(B) returns you back to the previous record. Press the POWER/MUTE
BUTTON (A) to return to the USER MENU.
SMART-CAL
From the working display, press & hold the POWER/MUTE BUTTON
(A) for 2-3 seconds. The display will read SMART CAL communicating.
Place the instrument into the cradle on the left side of the Smart-Cal
Calibration Station. Attach the tubing from the station to the instrument.
Press & release the CALIBRATE button on the Smart -Cal and calibration
will begin automatically. If successful, CALIBRATION PASSED will show
on display. If unsuccessful, CALIBRATION FAILED will show.
Let the instrument clear and repeat the calibration process. If the
instrument will not pass, remove the instrument from service. Consult
the factory in the event of any failure.