SensAlarm
Flex
–
Universal Gas Detection System
Sensidyne Document No. 360-0257-01 (Rev A)
P a g e
39 | 82
4.3.3.3.
Apply the span gas and the Gas Monitor will begin calibration automatically. If the
calibration gas does not begin automatically, check the tubing connection and that
the cylinder regulator is on.
4.3.3.4.
Once the Span Calibration is completed successfully, the following screen will be
displayed.
NOTE: If ENTER is not selected, the Gas Monitor reverts to the Normal Operation Main Display
after 5 minutes, and Normal Operations re-activates the 4
– 20 mA output.
4.3.3.5.
If Span Calibration fails, the following screen is displayed. Ensure adequate time has
passed for sensor stabilization, and that the calibration equipment is properly
connected with proper flow before reattempting the Span Calibration.
NOTE: If Span Calibration continues to fail, refer to the Troubleshooting Section in this manual. If
unresolved, contact Sensidyne Customer Support for assistance.
4.4. Test-on-
Demand™ (ToD)
This section covers the following
Test-On-
Demand™ (ToD) Mode Adjustment
submenu functions.
Enable and disable the automatic testing mode
Set the date & time for testing
Set the number of days between tests
Set the cell intensity
Designate output & fault indicators