Sensidyne Document No. 360-0213-01 - Rev B
© 2018 Sensidyne, LP
Page 15
Gilian Gilibrator
®
3 Dry Cell Calibrator – Operation Manual
3.6.
Connecting the Sampling Source
The sampling source to be calibrated may be either a positive pressure device
(exhausts air) or a negative pressure device (i.e. suction pump). Tubing must be
selected based on the flow range and fittings. The calibration base unit comes standard
with a 3/8 inch (OD) fitting. Step-down adapters are provided with the kit, (1) 3/8” to
1/4” adapter, and (1) 1/4” to 1/8” adapter.
Note: If sampling source will be used with a filter media, ensure that the media
is placed in between the sampling source and the calibrator to account for
backpressure. Failure to do so will alter the flow rates of sample.