Gilian Gilibrator 3

®

 Dry Cell Calibrator – Operation Manual 

 

3.5.  

System Set Up 

 

This  section  describes  the  steps  necessary  to  set  up  the  Gilibrator  3  Dry  Cell 

Calibration  System.  This  includes  initial  setup,  mounting  the  Flow  Cell  Assembly, 

connecting the tubing, and setting up the sampling source. Figure 3.6 shows how a 

complete Gilibrator 3 Dry Cell Calibration System may be configured. 

 

Prior to setup make certain you have properly connected the battery charger to the 

power port on the backside of the Base Control Unit and to an appropriate AC wall 

outlet. If the charger/adapter is properly connected, the Charge Indicator should light 

up. If you plan to use the Gilibrator 3 in the field, make certain the unit is fully charged 

before operating the unit. If you plan to use the unit in the lab/office (i.e., near an AC 

wall outlet), you can continue with setup and operation immediately. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 3.6 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for Gilian Gilibrator STABLFLOW

Page 1: ...ual This Manual Covers All Gilibrator 3 Kit Models Sensidyne Document No 360 0213 01 Rev B 1000 112TH Circle N Suite 100 St Petersburg FL 33716 USA 800 451 9444 1 727 530 3602 www Sensidyne com info S...

Page 2: ......

Page 3: ...or 3 Dry Cell Calibrator Operation Manual Gilibrator 3 Identifiers A LCD Display B Status and Notification LED C Ambient Light Sensor and Charge Indicator LED D SD Card Port E Power Switch F Power Por...

Page 4: ...d manufactured and calibrated in accordance with standards ISO 9001 2015 ISO IEC 17025 2017 ISO IEC 80079 34 ATEX Directive 2014 34 EU and IECEx where applicable Through ongoing review of our designs...

Page 5: ...uct s operating specifications or 2 is not properly maintained in accordance with its maintenance manual or specifications or 3 has been repaired or modified by persons other than authorized Sensidyne...

Page 6: ...brator 3 Dry Cell Calibration Kit Descriptions 8 SECTION THREE Set Up 9 3 1 System Components 9 3 2 Base Unit Components 10 3 3 Flow Cell Components 12 3 4 Preparation 13 3 5 System Set Up 14 3 6 Conn...

Page 7: ...ed without the prior written permission of Sensidyne LP Trademark Notice Sensidyne the Sensidyne logo Gilian and Gilibrator names and logos are registered trademarks of Sensidyne LP Other trademarks a...

Page 8: ...STORAGE OF THIS PRODUCT BY THE USER PERSON FIRM ENTITY CORPORATION OR PARTY NOT ADHERING TO THE INSTRUCTIONS AND WARNINGS IN THIS MANUAL OR OTHERWISE PROVIDED BY THE SELLER OR FROM NOT ADHERING TO ALL...

Page 9: ...r draw liquids of any type into the unit Pump failure faulting or user injury may result Do Not operate this unit with corrosive gasses or gasses that condensate Do Not pressurize the calibrator The G...

Page 10: ...y pack Do not connect the and terminals with metals such as wire Do not carry or store the battery pack with metal objects such as wire necklace or hairpins If the battery pack is short circuited exce...

Page 11: ...ications Approvals and Compliances The Gilibrator 3 is EN 61010 1 CE RoHS and EMC compliant The Gilibrator 3 contains an internal battery which has been approved for shipping and transport per UN DOT...

Page 12: ...tor 3 Dry Cell Calibration System include an easy to read color touch screen LCD display multi screen user interface with drop down menus and screen keyboard input capabilities corrections for standar...

Page 13: ...ent accuracy determines the overall accuracy of the unit The Gilibrator 3 Dry Cell Calibration System consists of two elements the Dry Flow Cell Assembly and the Control Unit Base The function of the...

Page 14: ...Single Cell Kit 1 Choose 1 Dry Cell 910 1702 US R Gilibrator 3 Low Flow Dry Cell 910 1703 US R Gilibrator 3 Standard Flow Dry Cell 910 1704 US R Gilibrator 3 High Flow Dry Cell Yes Two Cell Kit 1 Cho...

Page 15: ...timing system This type of microprocessor used in conjunction with the built in software provides an extremely accurate method for calculating the flow rate parameters The Gilibrator 3 Dry Cell Calibr...

Page 16: ...Light Sensor is located adjacent to the lower right corner of the LCD Display The LED indicates the status of battery charging The sensor allows the LCD backlight to increase to full brightness when...

Page 17: ...cle The Inlet Fitting is located on the side of the base unit towards the top The inlet is used for makeup air when calibrating a device in suction mode The inlet fitting can also be used to connect a...

Page 18: ...it takes the puck to travel between a pair of calibrated detectors White LED lights can be seen through the Viewing Window on Dry Cell These lights will illuminate when connected to the base and the b...

Page 19: ...arrives assembled and has interchangeable Flow Cells that connect into the base unit Figure 3 5 IMPORTANT Before proceeding you MUST charge the battery to full capacity prior to using the calibrator T...

Page 20: ...a complete Gilibrator 3 Dry Cell Calibration System may be configured Prior to setup make certain you have properly connected the battery charger to the power port on the backside of the Base Control...

Page 21: ...ure device i e suction pump Tubing must be selected based on the flow range and fittings The calibration base unit comes standard with a 3 8 inch OD fitting Step down adapters are provided with the ki...

Page 22: ...he base is powered down or when flow cells are disconnected all air paths are bridged to allow free flow of air from the inlet to the outlet and the puck is allowed to drop to the bottom of the cell O...

Page 23: ...ell type with flow range Reports Screen Reports Screen displays most recent calibration reports by date allows for preview and export of reports and for deletion of single or all reports Settings Scre...

Page 24: ...ecting Reference Temperature and Pressure Touch the button on the Screen adjacent to the STP icon The Input Keyboard will appear Note User may change from F to C on Settings Display Page Use keyboard...

Page 25: ...n an averaging set Choose from one of four pre selected values or select to open a keyboard window and select a value from 3 to 20 Type the number ranging from 3 20 of samples you wish to have average...

Page 26: ...glish Spanish French Portuguese German Dutch Italian Chinese Polish Turk Selecting Screen Brightness Touch the slider along the graduated marks to manually adjust screen brightness Furthest right is B...

Page 27: ...me Screen and corresponding records Selecting the Ambient Pressure Units Touch the buttons adjacent the Pressure icon to select from four units of pressure in H2O mmHg kPa or hPa Note The Ambient Pres...

Page 28: ...Description Step 1 Enter the Hour 0 23 Touch Next button Step 2 Enter the Minute 0 59 Touch Next button Step 3 Enter the Month 0 12 Touch Next button Step 4 Enter the Day 1 31 Touch Next button Step 5...

Page 29: ...tton to initiate the test This may take up to 5 hours do not unplug or turn off the unit while test is in progress A Please Wait window will appear while the test is in progress The unit will need to...

Page 30: ...t hose barbs Press the OK button Turn the calibrator upside down Wait for the piston inside of the flow cell to fall completely reach the top of the flow cell while upside down Return the unit to upri...

Page 31: ...the button with the Firmware Update icon A Firmware Update instruction window will appear Press OK button to begin update Do not turn off unit until the update is complete Set Sleep Timer Touch the bu...

Page 32: ...ned with the SmartCal format on the GilAir Plus Pump Prepare To Ship Touch the button with the Ship Gilibrator icon A Prepare to ship instruction window will appear Disconnect USB Cables Disconnect Po...

Page 33: ...splayed in the top left corner of the Home Screen The time and date can be edited on the Settings Display page Home Screen Sampling Mode is displayed on the top right corner of the Home Screen Averagi...

Page 34: ...ays the number of samples within an averaging set Note The sample count window does not appear in Continuous Mode Home Screen Battery Life is displayed in the bottom left corner of the Home Screen The...

Page 35: ...he averaging sample count data set Note The Gilibrator may take several cycles to stabilize Home Screen Flow Rate displays the live flow rate measurement from the Flow Cell Reset Sample Set Touch the...

Page 36: ...ord detail Recently Used Pump Screen This screen will populate with recently used pump information If unpopulated or the flow source name is not present touch the button labeled Skip to continue to in...

Page 37: ...nter in the unique sample identification Touch the button labeled Next to advance to the Operator ID keyboard Pump Operator ID Keyboard Enter in the operator information Touch the button labeled Done...

Page 38: ...sv If a post cal record is created and exported to the SD card the files will also appear in the same sub folder The post cal files will be denoted with a _ behind the sample ID ex D REPORTS Test 1 Te...

Page 39: ...port line highlighted touch the Delete button on the display screen Delete record window appears Touch the YES button to delete the single report Preview a Report With the report line highlighted touc...

Page 40: ...A Please Wait window will appear during data transfer Do not remove the SD card unitl transfer is complete If the file already exists on the SD Card a window will appear asking Overwrite Existing File...

Page 41: ...rom the unit by attaching it to a null model cable and using a terminal program like PuTTY The DB9 is a standard RS232 port and sends data in the familiar Gilibrator2 format at 2400baud 8N1 A header i...

Page 42: ...ollowing procedures to keep the unit in proper working order 1 Disconnect all cables from the Base Control Unit 2 Place black caps on both inlet and outlet fittings of Base Control Unit 3 Place black...

Page 43: ...uck inside the flow cell due to moisture aspiration it must be returned to Sensidyne for service Direct sunlight on the unit may trigger the internal optical sensors and cause the cell to be reset or...

Page 44: ...2018 Sensidyne LP Sensidyne Document No 360 0213 01 Rev B Page 38 Gilian Gilibrator 3 Dry Cell Calibrator Operation Manual NOTES...

Page 45: ...S A 800 451 9444 1 727 530 3602 fax 1 727 539 0550 www Sensidyne com info Sensidyne com Authorized EU Representative Schauenburg Electronic Technologies GmbH Weseler Str 35 45478 M lheim Ruhr Germany...

Page 46: ...1000 112th Circle N Suite 100 St Petersburg FL 33716 USA 800 451 9444 1 727 530 3602 www Sensidyne com info Sensidyne com...

Reviews: