Notice for User

 

 

 

 

 

 

Please  use  the  instrument  in  the  range  of  the  applicable  temperature, 
humidity  and  pressure  that  are  appropriate  for  the  specification  of  the 
product. Using the instrument beyond this range may cause malfunction 
or glitch of the instrument.

 

Gas  concentration  measurement  value  by  the  sensor  or  the  instrument 
can vary according to the environment at site (temperature, pressure and 
humidity). Therefore the calibration of the instrument should be performed 

at  the  same  or  similar  environment  as  that  of  the  instrument  use 
(temperature, pressure and humidity).

 

If temperature changes sharply during use of the instrument (for instance, 
using  the  instrument  at  places  of  far  different  temperatures  between 
indoor and outdoor), the value of the measured gas concentration can be 
changed  suddenly.  Please  use  it  after  the  gas  concentration  value  is 
stabilized.

 

Severe vibration or shock to the instrument may cause the sudden change 

of value of the measured gas concentration. Please use it after the value 
of gas concentration is stabilized. Excessive shock to the unit can lead to 
trouble of the sensor or the instrument.

 

Summary of Contents for SI-100C

Page 1: ...SENKO CO LTD SI 100C Operating Manual...

Page 2: ...g instruments are provided with a one year warranty This warranty covers only defective parts or workmanship in normal use and service Instruments which fail to function due to factory defect Within o...

Page 3: ...3 Copyright SENKO CO LTD Table of contents Warning 2 1 Introduction 4 2 Specification 4 3 Sensor specification 5 4 Electrical wiring 5 5 Calibration 6 6 Notice for user 8 7 Certification 9...

Page 4: ...ng Diffusion Type Measuring Gas Combustible Gas Toxic Gas Oxygen Measuring Method Catalytic IR Electrochemical cell Galvanic Accuracy 3 Full Scale Alarm 1st alarm 2nd alarm Output Signal Analog Output...

Page 5: ...20PPM 2PPM 5PPM H2 0 1000PPM 100PPM 500PPM H2S 0 200PPM 10PPM 15PPM NH3 0 100PPM 25PPM 35PPM LEL CH4 0 100 LEL 20 LEL 40 LEL Electrical Wiring PIN 3 Wire Color 1 Power 24V DC RED 2 4 20mA Signal Outpu...

Page 6: ...n it finished S Span Calibration Entry to Span Calibration It shows RED light on the LED marked Z when you supplied power to the instrument At this time press button once to move to Span Calibration m...

Page 7: ...ensor s output going down as time goes by measurement error will increase Calibration is crucial for the sensor performance and persons relying on this product for their safety WARNING Do not Fresh Ai...

Page 8: ...should be performed at the same or similar environment as that of the instrument use temperature pressure and humidity If temperature changes sharply during use of the instrument for instance using t...

Page 9: ...9 Copyright SENKO CO LTD Certification...

Page 10: ...10 Copyright SENKO CO LTD...

Page 11: ...11 Copyright SENKO CO LTD...

Page 12: ...12 Copyright SENKO CO LTD...

Page 13: ...13 Copyright SENKO CO LTD...

Reviews: