background image

 

 

Copyright SENKO CO., LTD. 

 

Calibration

 

 

Z: Zero Calibration

 

-Entry to Zero Calibration-

 

It  shows  RED  light  on  the  LED  marked  Z  when  you  supplied  power  to  the 
instrument.

 

-Execution of Zero Calibration-

 

Press the button for more than 3 seconds to execute zero Calibration at the state 
of lighting on the LED marked Z.

 

-Completion of Zero Calibration-

 

Zero Calibration proceeds for 3 seconds and will be completed. Processing time of 
Zero calibration may vary depending on each gas sensor applied. LED marked Z 
blinks during the process and will be lighted from Z to S repeatedly when it finished.

 

 

S: Span Calibration

 

-Entry to Span Calibration-

 

It  shows  RED  light  on  the  LED  marked  Z  when  you  supplied  power  to  the 
instrument. At this time, press  button once  to move to Span Calibration mode. 
You will be seen the AMBER light on the LED marked S.

 

It’s shifted from Zero(

RED

) to Span(

AMBER

) repeatedly by pressing the button.

 

-Execution of Span Calibration-

 

Prepare and supply the Standard gas into the gas sensor part of the instrument 
and then press the button for more than 3 seconds to execute Span Calibration at 
the state of lighting on the LED marked S.

 

-Completion of Span Calibration-

 

Processing time of Span Calibration vary depending on each sensor applied. LED 
marked  S  blinks  during  the  process  and  will  be  lighted  from  S  to  Z  repeatedly   
when it finished.

 

 

Summary of Contents for SI-100C

Page 1: ...SENKO CO LTD SI 100C Operating Manual...

Page 2: ...g instruments are provided with a one year warranty This warranty covers only defective parts or workmanship in normal use and service Instruments which fail to function due to factory defect Within o...

Page 3: ...3 Copyright SENKO CO LTD Table of contents Warning 2 1 Introduction 4 2 Specification 4 3 Sensor specification 5 4 Electrical wiring 5 5 Calibration 6 6 Notice for user 8 7 Certification 9...

Page 4: ...ng Diffusion Type Measuring Gas Combustible Gas Toxic Gas Oxygen Measuring Method Catalytic IR Electrochemical cell Galvanic Accuracy 3 Full Scale Alarm 1st alarm 2nd alarm Output Signal Analog Output...

Page 5: ...20PPM 2PPM 5PPM H2 0 1000PPM 100PPM 500PPM H2S 0 200PPM 10PPM 15PPM NH3 0 100PPM 25PPM 35PPM LEL CH4 0 100 LEL 20 LEL 40 LEL Electrical Wiring PIN 3 Wire Color 1 Power 24V DC RED 2 4 20mA Signal Outpu...

Page 6: ...n it finished S Span Calibration Entry to Span Calibration It shows RED light on the LED marked Z when you supplied power to the instrument At this time press button once to move to Span Calibration m...

Page 7: ...ensor s output going down as time goes by measurement error will increase Calibration is crucial for the sensor performance and persons relying on this product for their safety WARNING Do not Fresh Ai...

Page 8: ...should be performed at the same or similar environment as that of the instrument use temperature pressure and humidity If temperature changes sharply during use of the instrument for instance using t...

Page 9: ...9 Copyright SENKO CO LTD Certification...

Page 10: ...10 Copyright SENKO CO LTD...

Page 11: ...11 Copyright SENKO CO LTD...

Page 12: ...12 Copyright SENKO CO LTD...

Page 13: ...13 Copyright SENKO CO LTD...

Reviews: