background image

MGT Docking Station version II

13

7. Using Software

7.1 Overview of Software

Detector User ID

Calibration interval: 0~365 days

Low / High Alarm Setting

- H2S : 0 ~ 100.0

- CO  : 0 ~ 300

- O2  : 0 ~ 30.0

- LEL  : 0 ~ 100.0

Bump interval: 0~365 days

MGT firmware

STEL / TWA Alarm Setting

- H2S : 0 ~ 100.0

- CO  : 0 ~ 300

TWA Method

- ACGIH : Accumulation Average 

- OSHA : Movement Average

TWA

- Default : 8 hours 

average

STEL 

- Default : 15 minutes average

Safe Display : In the safe area, SAFE will displayed. 

Dock Lock : Lock the bump / calibration

Maintenance Notification

Latching Alarms : Until the users push the button, 

the alarm will be activated continuously. 

Auto Zero : When the device is turned on, the 

zero calibration will be activated. 

LEL by volume  :  100%LEL Standard

Off Lock : Prevent users from turning off the 

device.

To connect with USB

To connect with IR Link

Cal. Gas

Concentration

Upgrade

Figure 6

Enable/Disable

Cal./Bump Test

Serial Number

Firmware Version

Maximum Bump Time

(20 ~ 120sec)

Save power Mode 

(1 ~ 3600sec)

Bump result decision

- Beep

- Concentration

- Beep + Concentration

Docking 

Station 

firmware 

upgrade

Location

Gas Expiration 

date on cylinder

Read/Write

Cylinder No.

Figure 5

Summary of Contents for MGT Version II

Page 1: ...MGT Docking Station Version II User s Manual MGT Docking Station version II www senkocanada com sales senkocanada com...

Page 2: ...Inlet 2 5 Installation of Gas Cylinder 3 Testing 3 1 Calibration 3 2 Bump Test 4 Trouble Shooting 5 Docking Station II Manager Software 5 1 Software Installation 6 Connecting to the MGT Docking Statio...

Page 3: ...and devices Severe vibration or shock to the device may cause a sudden reading change and cause the device malfunction Do not leave or use the device at the watery place or at where there is any liqui...

Page 4: ...s It also maintains all the important event logs that demonstrate users are working within company requirements 4 1 2 Instructions and Parts 1 1 Description a b c d e f g h i j k l m n o p q No Descri...

Page 5: ...x 43 7 x 21 3cm 20 66 x 17 20 x 8 40 Weight Without cylinder 11 0kg Temperature 41 to 104 F 5 to 40 C Warranty Full 2 years Battery Type Rechargeable Lithium Ion Battery Life Up to 1 000 Bump Tests L...

Page 6: ...LEDs and meanings Each Bay has six blue LEDs see Figure 1 which indicates the process status Each of them will turn on by order from left to right while testing After testing is finished make sure all...

Page 7: ...straps and close the cylinder compartment Check the pressure gauge on the front panel Re install thumbscrews After installing a gas cylinder ensure USB is attached 3 Testing 3 1 Calibration Calibratio...

Page 8: ...therwise the third blue Bay LED illumination indicates if the MGT has failed the test see Figure 3 Figure 3 MGT has passed the test six LEDs MGT has failed the test only 3rd LED Bay locking mechanism...

Page 9: ...devices internal configurations The software can adjust and make changes to the Docking Station and the MGT units using SENKO IR Link and or partly USB Disk 5 1 Software Installation After downloadin...

Page 10: ...MGT Docking Station version II 10 The Installation will be ready to begin Click Install...

Page 11: ...ote The IR Link connection works only when it is placed into the Bay 1 USB drive must be installed to store the test results Do not open the software if IR Link is not plugged to PC When the Docking M...

Page 12: ...Link starts to flash rapidly After few seconds Read complete message will be displayed and current settings of Docking Station II will be showed on the interface Figure 4 Note If Send timeOut message...

Page 13: ...the bump calibration Maintenance Notification Latching Alarms Until the users push the button the alarm will be activated continuously Auto Zero When the device is turned on the zero calibration will...

Page 14: ...ftware 7 1 Overview of Software MGT Docking Station version II Figure 7 Load the data Refresh Click serial number Select the address Calibrated by Approved by Print All the information will be shown a...

Page 15: ...plays current setting of the Docking Station II and will discard any unsaved changes in the Tab Calibrate If the box is not selected calibration process will not proceed by pressing Calibration button...

Page 16: ...ed and alarm occurs alarm will not stop until resetting the unit push button If not selected and alarm occurs alarm will stop once present gas concentration goes below the alarm set points Auto Zero T...

Page 17: ...vents file in USB Disk Click Refresh button to view and download the latest logs After clicking Refresh button all logs will be available to view see Figure 8 To download the logs follow the below Plu...

Page 18: ...all data will be available to view see Figure 9 To print latest calibration certificate follow the below Plug the IR Link into a USB port on PC and place into Bay Unit 1 Double click and open the inst...

Page 19: ...or modify the product or any other cause of damage beyond the range of the intended use including damage by fire lightening water damage or other hazard voids liability of the manufacturer In the even...

Reviews: