Step 3:
continued...
Most likely the Reader Address, as shown in Figure 10, will be ‘1’.
This feature becomes relevant only when you have more than one
reader. Please note the reader address for sensor calibration.
Figure 10: Results of Reader Search
Step 3:
continued...
Once the physical setup is complete, determine the address of the
wireless interrogation unit by selecting “Search for Readers” under
the Readers menu as shown in Figure 9.
Figure 9: Search for Readers
Step 3:
continued...
Select “Calibrate” from the Sensors Menu to obtain the Calibration
Screen. Sensors are identified by model numbers that reflect
nominal the frequency of each sensor. Since the six sensor
frequencies can be interrogated by three antennas, the Calibration
Screen allows the user to select from 18 antenna and sensor
combinations. However, as shown in Figure 11, the FCC compliant
version allows the user to only use Antenna 1 (antenna ports 2 and
3 are deactivated). This is a regulatory requirement. The user can
select up to 6 sensors to be calibrated from the list of available
sensors in the Calibration Screen. Alternately, the CE compliant
version allows the user to only select sensors that operate in the
433MHz band.
Figure 11: Calibration Screen - with FCC version