USER MANUAL – Z-SG
13
Use the software “Easy Setup” (download from
) for Configure and
Calibrate the Z-SG/Z-SG-L.
10.2.
CALIBRATION WITH A STANDARD WEIGHT
WARNING
Gross weight (tare + Standard weight) must not to exceed load cell end scale, to avoid
serious damage to the cell.
1) Power off the module.
2) Switch Dip-Switch SW2-
1 as desired: “OFF”=digital input enabled, digital output disabled;
“ON”=digital input disabled, digital output enabled
3) Switch Dip-Switches SW2-2 and SW2-3 as desired: see Dip-Switches table
4) Switch Dip-Switches SW2-4
to “OFF” and SW2-5 to “ON”
5) Switch Dip-Switches SW2-
6 to “ON”, SW2-7 to “ON”, SW2-8 to “ON”
6) Power ON the module
7) Write sensitivity value in reg. 40044-40045
8) Write the Standard Weight value in reg. 40048-40049
9) Reset the module (write 0xABAC=43948 in reg.40068)
New sensitivity and Standard weight are saved in Z-SG/Z-SG-L module.
10) Put the tare on the balance
11) Save the tare value in EEPROM memory (write 0xC2FA=49914 in reg.40068)
12) Put the known weight on the tare
13) Save the known weight in EEPROM memory (write 0xC60C=50700 in reg.40068)
14) Switch Dip-switch SW2-1 OFF SW2-2 ON SW2-3 ON SW2-4 OFF SW2-5 ON SW2-6 ON
SW2-7 ON SW2-8 ON
10.3.
CALIBRATION WITHOUT A STANDARD WEIGHT
1) Power off the module before configuring it by Dip-Switches to avoid serious damage due to
electrostatic discharges.