background image

4/4

TECHNICAL SPECIFICATIONS

STANDARDS

EN61000-6-4

 

Electromagnetic emissions, industrial environment.

EN61000-6-2

 

Electromagnetic immunity, industrial environment.

EN61010-1    

Safety

INSULATION

ENVIRONMENTAL  

CONDITIONS

Temperature

:               -20 

÷

 + 65°C

 

Humidity

:                     10% 

÷ 

90% non condensing.

Altitude:                       

up to 2000 m above sea level

 

Storage temperature:

  -40 

÷ 

+ 85°

 

Protection degree:       

IP20.

ASSEMBLY

IEC EN60715, 35mm DIN rail through terminals.

CONNECTIONS

5-way removable screw terminals, pitch 5 mm

POWER SUPPLY

Through PC USB port.

SERIAL 

COMMUNICATION

RS232 (DB9-M), RS485 (5 poles), TTL (RJ10), USB 1.0, 1.1, 2.0

BAUDRATE

1200 bps,115200 bps.

WARNING

the maximum working voltage between any termi

-

nal and ground must be less than 50Vac/75Vdc

Power

Supply

Input

Output

1500 V

~

RS485

USB

In/Out

RS232

TTL

INSTALLATION REGULATIONS

As regards the S117P1 instrument, there are no particular precautions to follow, except that of using a shielded 

cable for the RS232 and RS485 lines for long connections or connections in noisy environments.

TERMINAL POSITION AND DESCRIPTION

The drawing below shows the functions of the removable screw connector on the side of the S117P1:

PIN INDEX

DESCRIPTION

1

Power Supply Ground

2

Max: + 12Vdc @ 100 mA

3

A

4

B

5

Modbus Ground

CONTACT INFORMATION

Technical support

[email protected]

Product information

[email protected]

This document is the property of SENECA srl. Copies and reproduction are prohibited unless authorised. The con-

tent of this document corresponds to the described products and technologies. Stated data may be modified or 

supplemented for technical and/or sales purposes.