![Sendoline S1-A Operating Instructions Manual Download Page 33](http://html.mh-extra.com/html/sendoline/s1-a/s1-a_operating-instructions-manual_1239356033.webp)
6 Follow-up
Operating Instructions S1-E / S1-A
30
65 67 460 D3643
.
201.01.02
.
02
6
Follow-up
6.1
After each treatment session
Immediately after treatment
NOTICE! Condition immediately, or at the latest, one hour after
treatment.
Sendoline spray
NOTICE! Only use W&H Service Oil F1, MD-400.
Instruments _ File
✔ Wear appropriate protective clothing.
1. Remove the file with tweezers.
2. Pre-disinfect directly at the treatment center [ → 32].
3. Detach the instrument from the drive.
4. Transport the instrument to the hygiene room in a suitable transport
container.
5. Conduct automatic conditioning [ → 33]. Manual conditioning [ →
36] is possible in exceptional cases if the national/local regulations
are followed.
6567460.book Seite 30 Donnerstag, 8. September 2016 12:16 12