![Sendoline S1-A Operating Instructions Manual Download Page 23](http://html.mh-extra.com/html/sendoline/s1-a/s1-a_operating-instructions-manual_1239356023.webp)
5 Operation
Operating Instructions S1-E / S1-A
20
65 67 460 D3643
.
201.01.02
.
02
5
Operation
Fatigue fracture
NOTICE! Use only faultless files in order to prevent fatigue fractures of
the files as far as possible.
Zurückziehen der Wange _ Instrument
CAUTION! Do not pull the patient's cheek back with the contra-angle
handpiece while the motor is running. This would actuate the
pushbutton, thus creating a risk of burning the patient's oral mucosa.
Maximum motor speed for S1-A
CAUTION! Use the S1-A only with an air-driven motor at a maximum
speed of 22,000 rpm.
Observing manufacturer documentation
IMPORTANT: For application details and operating data of the different
preparation tools, please refer to the information provided by the
manufacturer.
5.1
Replacing the instrument
Motor at standstill
CAUTION! The instrument should only be fitted or removed when
the motor is at standstill.
6567460.book Seite 20 Donnerstag, 8. September 2016 12:16 12