
11.
ADDENDUM
11.1 MULTI-POINT CALIBRATION
11.1.1 Introduction
If the distribution of film weight on your calibration panel is not
uniform it may be possible to achieve a more accurate calibration of
the SI9500 by calibrating at several points on the panel and taking
the average. Special software has now been produced to achieve
this and is available in units with software version 1.17 and higher.
The version can be displayed by pressing CLEAR and END
simultaneously while in Measure mode.
11.1.2 Setting for Multi-Point Calibration
Refer to appendix 1 of your manual sections A1.2, A1.5 and A1.6.
Switch position 4 should be closed. This will enable the multi-point
calibration facility.
11.1.3 Operation
1.
Press CALIBRATE to enter calibration mode.
2.
Select lacquer type.
3.
Place probe on sample and press ENTER.
4.
Repeat stage 3 for as many points as required.
5.
Press END or select the next lacquer type.
Summary of Contents for SI9500
Page 11: ...OPERATION KEYS NUMERIC KEYPAD ...