SellEton Scales PS-IN202SS User Manual Download Page 11

PS-IN202SS Indicator 

- 10 -

V2015.05 

Chapter 3

Calibration 

The indicator is calibrated by following the procedures embedded in P11 (Zero) and P12 
(Span) of the Setup Menu. Each procedure enters a value into the indicator's non-volatile 
memory: 
P11 is the zero value (deadweight) 
P12 is the span value (test weight). 

The minimum test weight that can be used is 1% of the full-scale capacity. After the two 
calibration procedures are executed successfully, a record should be made of both 
calibration values in the Calibration Table below using the P 18 View procedure. 

NOTE: 

This 

chapter assumes th

at the indicator is in Setup (“P”) Menu mode. 

Calibration Table 

Indicator 

Zero Calibration 
Span Calibration 
Serial Number 

3.1 Zero Calibration 

a. While in the Setup mode, scroll to "P 11", then scroll down once using the 

[ZERO]

 key to

enter the zero calibration menu. The display will momentarily show "C 0" followed by a value.
This value is the internal A/D count and is useful when trying to troubleshoot setup problems.

b. Ensure there are no test weights on the platform and press the 

[ZERO]

 key again to zero

out the displayed value.

c. Press the 

[NET/GROSS]

 key to save the zero point value. The display will show "End C0"

momentarily then revert back to P11. Proceed to the P12 span calibration to complete the
indicator calibration.

3.2 Span Calibration 

a. While in the Setup mode, scroll to "P 12", then scroll down once using the 

[ZERO]

 key to

enter span calibration menu.

b. The display will momentarily show "C 1" for the span calibration, followed by a value
with one flashing digit. This value will be zero with the Decimal Point parameter as selected
in P10. Place the test weight on the weighing mechanism.

c. Use the four direction keys to adjust the displayed value to the actual test weight value.
Increment the flashing digit by pressing the 

[UNITS]

 key. Decrement the flashing digit by

pressing the 

[ZERO]

 key. Pressing the 

[PRINT]

 key or the 

[TARE]

 key will change the

position of the flashing digit.

d. After setting the exact value, press the 

[NET/GROSS]

 key to save the value.

Summary of Contents for PS-IN202SS

Page 1: ...Version 2013 08 PS IN202SS Weight Indicator User s Manual...

Page 2: ...R 2 CONFIGURATION 2 1 SET UP P MENU 2 2 Setup Menu Chart 2 3 Exiting the setup menu 2 4 USER S MENU 2 5 User Menu Chart 2 6 Setup Menu Descriptions 2 7 User Menu Descriptions CHAPTER 3 CALIBRATION 3 1...

Page 3: ...PS IN202SS Indicator 2 V2015 05 Front View of the Indicator FIG1 shows the front view of the indicator FIG 1...

Page 4: ...as shown below in FIG 2 Assemble the cables with the pin assignments shown in FIG 2 FIG 2 1 3 CONNECTION TO A PRINTER OR COMPUTER The indicator contains a standard full duplex RS 232 serial port desi...

Page 5: ...own is the weighing mode c Switch on the indicator The indicator will display P 1 to indicate that it is in Setup P menu mode d Choose the channel you are working on The default channel is 1 You can p...

Page 6: ...Calibration position Up is the calibration mode Down is the weighing mode c Switch on the indicator The display will go through a digit check then go into Normal Operating mode All front panel keys w...

Page 7: ...Chart d To move to the selection level press the ZERO right key once The current saved selection is shown e To view the available selections for the current S heading use the TARE down or PRINT up ke...

Page 8: ...m value varies depending on the local regulations 1d 3d 5d 10d P8 Digital Filter Averages weight readings to produce higher stability The higher the filter setting the greater the stability but the sl...

Page 9: ...op bit 7E 7 data bits with even parity bit and one stop bit 7n 7 data bits with no parity bit and two stop bits 8n 7O 7E 7n S3 Mode of Serial Transmission Selects when data will be sent out of the ser...

Page 10: ...int the header information Valid only when S6 is set to 1 0 Do NOT Print Header 1 Print Header 0 1 S12 Back light set Set back light mode OFF ON and Auto when the display is idle the back light will t...

Page 11: ...nternal A D count and is useful when trying to troubleshoot setup problems b Ensure there are no test weights on the platform and press the ZERO key again to zero out the displayed value c Press the N...

Page 12: ...libration menu b The display will momentarily show CAL 0 followed by a value This value is the zero calibration value and should be recorded in the table below Press any key to continue c The display...

Page 13: ...he displayed value to the span calibration value c After setting the exact value press the NET GROSS key to save the value d If the entered value is greater than zero the display will show E CAL 1 mom...

Page 14: ...isplaying the gross weight T TARE The tare weight has been established in the system lb kg oz PCS lb kg oz PCS Indicates the unit of the displayed weight PCS pieces STABLE This light is illuminated wh...

Page 15: ...the weight shown on the display 4 5 Tare Function To weigh an item in a container the weight of that container must first be subtracted from the overall weight to obtain an accurate weight reading Thi...

Page 16: ...ve until all the weight values are added Press and hold ACCU for 2 seconds to enter the accumulation review mode It will show the first value added Press ACCU once and it will move on to the second va...

Page 17: ...PS IN202SS Indicator 16 V2015 05...

Reviews: