CALIBRATION PROCEDURE
KEY
Function
ON/OFF
Exit
SET
Confirm
ZERO
Move Up
UNITS
Move Down
HOLD
Move Left
TOTAL
Move Right
1. Turn on the scale by holding ON/OFF for 2 seconds.
2. Press ON/OFF
and SET
together to access the setup menu.
3. If done correctly, the display should now show
C01
.
4. Press SET
to access the C1 channel. The display should show [
C1
#].
5. Press ZERO to change the value of C1 to [
C1 2
]. (The value of 2 sets the scale
6. up in pounds.)
7. Press SET
to set the value. The display will now show
C02
.
8. Press SET
to access the C2 channel. The display should show [
C2
#].
9. Press ZERO
to change the setting to the decimal places desired.
(The C2 channel is used to adjust the decimal point on the scale. A value of 1 means
there is one digit behind the decimal point.)
10. Press SET
to set the value. The display will now show
C03
.
11. Press SET
to access the C3 channel. The display should show [
C3
#].
12. Press ZERO
to cycle through the values until the desired graduation appears.
(The C3 channel adjusts the divisions on the scale. A value of 1 selected and C2 set
to 1, the scale will read in 0.1 lb. increments.)
13. Press SET
to set the value. The display will now show
C04
.
14. Press SET
to access the C4 channel. The display will show [######].
15. Enter in the maximum capacity you want to use for this scale by using HOLD and
TOTAL to move the cursor left and right, and ZERO
and
UNITS and to move
the values up and down. (The C4 channel is used to enter in the max capacity of the
scale; Make sure this doesn’t exceed the max capacity of the scale; Max capacity
divided by the increment set in C02 and C03 above cannot exceed 5000)
16. Press SET
to set the value. The display will now show
C05
.
17. Press SET
to access the C5 channel. The display should show [
C5 0
].
18. The C5 channel calibrates zero on the scale. Make sure the scale is empty.
19. Press ZERO
to change the value to 1.
20. Press SET
. The display will count down from 10-1 while the scale is
calibrating zero. When the display shows 0 the zero calibration is complete.
21. Press SET
to continue. The display will now show
C06
.
22. Press SET
to access the C06 channel. The display will show [
C6 0
].