6 Description and usage
The autoclave ST Dry PV 12 litres is an equipment useful for sterilizing processes in
clinics departments and hospitals, of instruments and devices wraped and unwraped,
prions and drying. The software programme provides auto diagnose, protections
against over heating or over pressurizing, and ensure reliability of the sterilization
results. They have a condensed water collection box which keeps the equipment safe.
This autoclave uses the advanced technology of 3 times fractionated and pulsatory
driying vacuum. It enables the hot steam to penetrate into the porous material.
6.1 Location of elements
5 Technical specifications
Chamber inner dimensions (diameter x depth): ....................
200X358mm
Power supply: ..............................................
according to nameplate
Sterilizing temperature/pressure: ........................
121ºC/(1.0-1.3) bar
Sterilizing temperature/pressure: ........................
134ºC/(2.0-2.4) bar
Power fuse: ........................................................................
F12A 250V
Water reservoir capacity: ......................................................
3.2 litres
Operation temperature range: ..............................................
0-40ºC
Water minimum level: ..........................................................
1L
Environmental conditions: .................................................
+5ºC to +40ºC
Relative humidity: ............................................................
35% to 75%
1 Door knob
2 Control panel
2
1