2)  Sample placing

Put the sample (cover is up) on stage A6 and fit it with clamp A7.  Turn B7, B8 stage knobs to put the sample 

into the light path (Fig. 4).

u

   

Two samples can be placed on the stage together

3)  Focusing (Fig. 5)

Put 10x objective into light path, turn coarse 

focusing knob B13, observe from right eyepeice 

with right eye, then turn slowly fine focusing 

knob B14 to make the image clear after finding image.  

u   

The tension adjustment ring B12 can adjust the tension of the coarse and fine focusing unit to prevent    

   

the stage from sliding down automatically and to improve the comfort of operation.  Rotate clockwise to  

   

decrease tension, rotate counter-clockwise to increase tension.

Don’t turn left and right coarse and fine focusing knobs in different directions with power at the same    

 

time, if so, the focusing system will be damaged. (Fig. 6)

4)  Interpupillary distance and diopter adjustment

a.  Interpupillary distance adjustment

  Interpupillary distance is different for everyone, so interpupillary 

  distance should be adjusted before using binocular microscope. 

  Please take the two eyepiece tubes to turn until the bright rings 

  observed by two eyes are in superposition entirely. (Fig. 7)

b. Diopter adjustment (Fig. 8)

  As focusing for binocular, user should observe right eyepiece with right eye and make the right eyepiece

  clear by focusing adjustment, then observe the left eyepiece at the same time, adjust the diopter ring of the    

  left eyepiece tube to make the image of the left eyepiece clear as same as the right eyepiece.

Fig. 4

Fig. 5

Fig. 6

7

Summary of Contents for Microlux IV

Page 1: ...Microlux IV LED Operations Manual...

Page 2: ...3433 Tree Court Industrial Blvd St Louis MO 63122 800 489 2282 314 968 2282 Email micro seilerinst com www seilerinst com...

Page 3: ...Contents 1 Parts Name 1 2 Specification 2 3 Installation 3 4 Operation 4 5 Installation Operation for Attachments 5 6 Maintenance 5 7 Troubleshooting 10 8 Outfits 11...

Page 4: ...r B4 Condenser Adjusting screw left right A8 Condenser A10 Filter B5 Field diaphragm revolver B6 Condenser thumb screw B7 B8 Stage moving knobs B9 Potentiometer A11 Conden set holder A6 Mechanical sta...

Page 5: ...er A12 Handle A13 Power Input A14 Fuse A15 Lamp Plate B1 Head thumb screw B2 Diopter adjuster B3 Aperture diaphragm lever B4 Condenser adjusting screw B5 Field diaphragm revolver B6 Condenser thumb sc...

Page 6: ...m X Y travel 55mm X 78mm 2 3 6 Focusing systems Coaxial coarse and fine focusing knobs Coarse stroke 25mm fine division 2 m Condenser up down range 25mm 2 3 7 Condenser Abbe condenser N A 1 25 Adjusta...

Page 7: ...lter Fig 2 3 Installation Please install the microscope as follows Installation Process see Figure 3 on page 5 1 Installing Lamp LED is factory installed 2 Installing binocular head Loosen B1 binocula...

Page 8: ...d Turn B10 condenser focusing screw and raise the condenser to working position 6 Installing blue filter or the other attachment If blue filter will be used the installation should be done before proc...

Page 9: ...clamp and put the sample on it turn B7 B8 stage knobs to put the sample into light path Turn A4 nosepiece put 10x objective into light path turn B13 B14 coarse and fine focus knobs Adjust A2 head to m...

Page 10: ...ase tension u Don t turn left and right coarse and fine focusing knobs in different directions with power at the same time if so the focusing system will be damaged Fig 6 4 Interpupillary distance and...

Page 11: ...move the image to the center of field d Put 40x objective into light path turn B5 field diaphragm revolver to make the field diaphragm image bigger than field diaphragm Please center it again if the i...

Page 12: ...ith hand Dust on lens should be cleaned by soft brush absorbent cotton or lens paper with a mixture of alcohol and ether proportion 1 4 6 1 2 Alcohol and ether are flammable please keep them away from...

Page 13: ...occurs please refer to the following chart for some common resolutions Trouble Power on but no light Brightness of view field isn t enough or uneven Cause Plug is unreliable Bulb is broken Fuse is br...

Page 14: ...adjusted correctly Diopter isn t adjusted correctly Trouble Brightness of view field isn t enough or is uneven Image isn t clear contrast or definition isn t enough One side of image is dark or image...

Page 15: ...3601 www seilerinst com Microscope Division Manager Dane Carlson ext 365 dcarlson seilerinst com Corporate Sales Representative Nicole Rasch ext 345 nrasch seilerinst com Technical Product Information...

Page 16: ......

Reviews: