20
Specifications
1-1 General Specifications
Display Count :
6000 count
Overrange Display :
OL or -OL
Conversion Rate :
3 times/second
Dimensions(W x H x D) :
60 x 220 x33 mm
Weight :
260g
Power requirement :
AAA size 1.5 x2
Battery Life :
150 hours Alkaline battery
Maximum Conductor Size :
20mm
LVD :
EN61010-1, EN61010-2-032, EN61010-2-033
EMC:
EN61326-1
Installation Category :
CAT.III. 600V, CAT.II. 1000V,
CAT.IV. 300V
CAT Application field
Ⅰ
The circuits not connected to mains.
Ⅱ
The circuits directly connected to Low-voltage
installation.
Ⅲ
The building installation.
Ⅳ
The source of the Low-voltage installation.
1-2 Environmental Conditions
Indoor Use.
Maximum operating altitude :
2000m (6562ft)
Operating temperature :
-10°C ~ 30°C,
≦
80%RH
30°C ~ 40°C,
≦
75%RH
40°C ~ 50°C,
≦
45%RH
Storage temperature :
-20 to +60°C, 0 to 80% RH (no batteries).
Temperature coefficient :
0.2 x (Specified accuracy) / °C, < 18°C, > 28°C
Pollution Degree :
2
Shock vibration :
MIL-PRF-28800F for A class 2 Instrument
Drop Protection :
4 Feet Drop to hardwood on concrete Floor