![SEFAM PolyLink Procedure For Installation And Usage Download Page 8](http://html.mh-extra.com/html/sefam/polylink/polylink_procedure-for-installation-and-usage_1233311008.webp)
8
Installation
PolyLink kit
Installation
Insertion of the PolyLink module in the S.Box device
Refer to the patient manual for the instructions concerning the installation of the S.Box device before
inserting the PolyLink module into the device, as described hereafter:
1.
Unplug the S.Box device from the power supply or from the electrical power grid.
2.
Remove the side cover or the humidifier from the device:
Press the button for unlocking the cover or the humidifier chamber and pull simultaneously
the cover or the chamber towards you by using the corresponding integrated handle.
In addition, if the S.Box is equipped with a humidifier, unlock the heater plate by pushing the
latch to the right and remove it.
3.
Raise the lever of the PolyLink module and slide the
module into the accessory location, making sure to
position the connector upwards. Once it is in place, lock
the module by pushing the lever down.
Figure 3 – Insertion of the module into the device
4.
Re-place the side cover or the humidifier into the device:
Reinsert the cover on the rails in the side part of the device and push it against the device until
you hear a "click".
If the S.Box is equipped with a humidifier, reinsert the heater plate on the rails
in the side part
of the device and push it against the device. Re-lock the heater plate by pushing the latch to
the left. Install the base of the humidifier chamber on the heater plate and push the chamber
against the device until you hear a "click".
5.
Plug the S.Box device into the power supply or electrical power grid.
Implementation of the effort/position kit and sensors
1.
Attach the support to the Thorax effort belt.
2.
With a thin tip like that of a pen, turn on the PolyLink effort/position kit by setting the switch to On
(sliding door to left) and check that the LED is lit in green, indicating that the internal battery is
charged.
If the LED is lit in orange or in red, recharge the internal battery according to the instructions in the
corresponding paragraph, page 9).
3.
Insert the PolyLink effort/position kit into its support
positioning it in the direction indicated by on the kit,
the side with symbols facing outward
(A).
4.
Install the two effort sensors on the thorax and
abdomen following the instructions of the user
manuals provided.
5.
Connect the cable of the thorax sensor (blue
connector) and the cable of the abdomen sensor
(yellow connector) to the ports of same colours of the
PolyLink effort/position kit
(B).
Figures 4 : Implementation of the effort/position
kit (A) and sensors (B)
A
B