EcoStar Info, EcoStar Auto
Accessing settings
9
Features available on the devices:
Feature
Feature
enabled
Feature
disabled
Default
Comfort Calibration
CC:1
(continuously or
during the ramp )
CC:0
CC:1
(during the
ramp )
External module connection
rS:1
rS:0
rS:1
Command on Inspiratory Flow Limitation (in
Auto-CPAP mode only)
FL:1
FL:0
FL:1
Auto ON
Ao:1
Ao:0
Ao:1
Ramp type
Default
Ramp feature
Time ramp
(tr)
Indexed ramp
(ir)
Indexed
ramp (ir)
Pneumatic configuration of the devices regarding the accessories used:
Display calibration pneumatique
Accessory used
Without humidifier
With GKH
2
O heated
humidifier
Tubing 1.80 m long 22 mm diameter (standard
configuration set by default)
CP:22
CP:HH
22
Tubing 1.80 m long 15 mm diameter
CP:15
CP:HH
15
Other (specific)
CP:SP
(specific calibration present)
CP:--
(specific calibration absent)
Accessing the menus
In order to restrict access of certain settings to the
medical team, a hidden key has been provided on the
device. It is located under the letters "Ec" of the name
"EcoStar
” and is detectable to the touch
as a hollow.
Moreover, the three buttons shown at the user
interface allow access to the device settings and are
used to change the value of certain parameters.
Description of a screen page
Each screen of the practitioner settings menu includes:
an upper section indicating the value of the
parameter displayed
a lower section with various symbols indicating the
device's operating mode or the nature of the
parameter
displayed
(see
table
entitled
"Description of symbols displayed on the screen").
The parameters accessible on the screen
when the machine is in standby or
operating mode are:
parameters of the device relating to
treatment
compliance counter data.