LECTRICAL SAFETY INSTRUCTIONS
As the appliance is not sealed, never immerse it in liquid. Never use it in a
flammable environment.
Never use the
Premio 10 moxa
during a thunderstorm.
Channeled
electrical discharges are
uncontrollable and dangerous for both the
appliance and the users
.
Be sure to only use the 110V/240V, 50/60 Hz medical power supply box
which supplies 5V between 1.5 and 3 A, or if not, a protected 5V external
battery with an isolating cover, which is not connected to the mains
whilst using Premio 10 moxa and does not supply more than 3 A.
Caution, it must be a medical power supply unit. Otherwise, you could be
exposed to severe risks.
The power supply box must always be accessible so that the device can
be disconnected from the mains.
Never leave the unconnected power cord outside of its box.
T
HERMAL SAFETY INSTRUCTIONS
When in use, some parts of the
Premio 10 moxa
become hot.
This means basic care is needed when handling:
-
do not use the
Premio 10 moxa
in the presence of flammable
gas (oxygen, alcohol or ether vapors, etc.)
-
Stop the device before cold cleaning
-
after cleaning, wait until the disinfectant has totally evaporated
before turning the appliance on
-
do not touch hot areas: metal parts of the diffuser and
Concentrator.
-
The temperature of the front part between the 2 symbols
can exceed 48°C, after 5 minutes of continuous operation.