PrimeTest 300
Operating
Instructions
Disconnect the black test lead from the unit.
Disconnect the Seaward mains cable, from the mains
inlet socket, ref fig. 2 (7).
Disconnect the IEC mains cable, from the IEC socket ref
fig 2 (8).
Disconnect the EUT mains cable, from the EUT socket,
ref fig. 1 (6).
Position the PrimeTest 300 face down and release the
captive screw in the battery compartment cover.
Remove the battery compartment cover.
Lift one end of the fuse out of the fuse holder with the
help of a flat bladed screwdriver.
Lift the defective fuse completely out of the fuse holder.
Insert a new fuse as described and specified by the text
on the battery compartment cover.
Ensure that the new fuse is seated and centred in the
fuse holder.
Relocate the battery cover over the battery compartment
and fasten in position with the battery cover captive
screw.
16.6 Service and Calibration.
To maintain the specified accuracy of the measurement
results, the instrument must be recalibrated at regular
intervals by either the manufacturer or an authorised
Seaward Service Agent. We recommend a recalibration
period of one year. If a period of greater than 1 year has
elapsed since the last calibration/validation a warning
message will appear on the display every time the unit is
powered up.