OPERATION
Data Logging
Select Set Up Logging from the sensor screen
If there are no files currently on the sensor you’ll see the
Set Up Logging button active under the Data Files tab as
well as in the upper menu. Once files have been started/
logged on the sensor they will be displayed under the Data
Files tab.
Setting Up Logging Window
Here you will name your data file and set up the recording
interval and duration of each logging phase. Select your
desired recording interval and duration for each phase,
Aqua4Plus will display the available memory at the bottom
of the window.
Click to switch between interval and continuous
data recording (supported sensors only) Select your
continuous rate from the drop down box here.
Duration can be set by either number of records
Or by setting a duration time
When set by number of records the time of the recording
phase will be displayed detailing how long that phase will
run. When set by time, the total number of records for that
phase will be displayed.
If you need to check settings or perform a calibration click
before proceeding with logging setup to switch to the
Settings and Calibration screen.
You may sync the sensor clock with the PC clock when
starting logging by clicking the slider here.
Check the Delayed Start box and enter the desired date/
time you would like logging to start. This is useful for
syncing data when setting up multiple sensors on a site.
Data will start logging at the set date/time rather than
immediately when Start is pressed.
Data file name defaults to Test File # and may be re-named
here.
The 3 previous Logging Schedules that were programmed
to a sensor will be listed under the Select Template drop
down menu. There you will also find pre-programmed
logging schedules such as 24 hour pump test, along with
any custom logging schedules saved by the user.