background image

tP101.V2  | issue:1   30/11/17

Original Language Version

© Jack sealey limited

6. 

TROUBLEShOOTING

problem

Cause

Solution

no indication on display

Bad battery contact

check battery contacts

inaccurate measurement

Wrong calibration

follow calibration procedure

the meter is working below minimum 

acceptable flow rate

i

ncrease the flow rate until an acceptable flow rate range 

has been achieved

r

educed or zero flow rate

turbine blocked

clean the turbine

t

he meter does not count but the flow 

rate is correct

incorrect installation of the card

reinstall the card

Possible electronic card problems

contact your stockist

Sealey Group, Kempson Way, Suffolk Business Park, Bury St Edmunds, Suffolk. IP32 7AR

 01284 757500 

 01284 703534 

 [email protected] 

 www.sealey.co.uk

Note

: it is our policy to continually improve products and as such we reserve the right to alter data, specifications and component parts without prior 

notice.

Important

: no liability is accepted for incorrect use of this product.

Warranty

: Guarantee is 12 months from purchase date, proof of which is required for any claim.

WEEE REGULATIONS

dispose of this product at the end of its working life in compliance with the eu directive on Waste electrical and electronic equipment 
(Weee). When the product is no longer required, it must be disposed of in an environmentally protective way. contact your local solid 
waste authority for recycling information.

BATTERY REMOvAL 

refer to section 5.2.2
under the Waste Batteries and Accumulators regulations 2009, Jack sealey ltd are required to inform potential purchasers of products 
containing batteries (as defined within these regulations), that they are registered with Valpak’s registered compliance scheme. Jack 
sealey ltd Batteries Producer registration number (BPrn) is BPrn00705.

ENvIRONMENT pROTECTION

recycle unwanted materials instead of disposing of them as waste. All tools, accessories and packaging should be sorted, taken to 
a recycling centre and disposed of in a manner which is compatible with the environment. When the product becomes completely 
unserviceable and requires disposal, drain any fluids (if applicable) into approved containers and dispose of the product and fluids 
according to local regulations.

Summary of Contents for TP101.V2

Page 1: ... liquids in accordance with local authority regulations 8 8 DO NOT exceed the maximum pressure of 50bar 8 8 DO NOT use the equipment near open flames 8 8 DO NOT smoke whilst using this equipment 8 8 DO NOT use for corrosive fluids 8 8 DO NOT dismantle tamper with or adapt the equipment for any purpose other than for which it is designed 8 8 DO NOT use the unit if it has been dropped or mishandled ...

Page 2: ... The meter features two buttons RESET and CAL which individually perform two main functions and together perform other secondary functions 4 4 1 1 RESET key is used to reset the Batch TOTAL and Reset TOTAL 4 4 1 2 CAL key is used to enter the calibration mode 4 4 1 3 Combination of RESET CAL keys is used to change the unit of measurement and other secondary functions 4 4 2 Turbine Assembly it has ...

Page 3: ...our combinations of units of measurement as shown in fig 7 4 8 3 Press CAL key for more than 2 seconds to store new settings The METER will return to Standby Mode see fig 8 Note No new calibration is required after changing the Unit of Measurement 4 9 Normal Dispensing Mode see fig 9 4 9 1 While the media is flowing through the meter Batch Total and Reset Total are displayed at the same time 4 9 2...

Page 4: ...the display page is showing the Reset TOTAL is shown press and hold the Reset key again until the Resettable TOTAL turns to zero see fig 17 and fig 18 4 11 1 4 Finally the page with the new Reset TOTAL is displayed see fig 19 4 12 Calibration 4 12 1 In standby mode press and hold the CAL key to see the current calibration factor Factory K Factor Factory set default factor It is equal to 1 indicate...

Page 5: ...isplayed 4 12 4 1 In field Calibration Sequence 1 Wait until the METER returns into Standby display shows TOTAL see fig 22 2 Press CAL key for more than 2 seconds The METER enters calibration mode and shows CAL The words FACT or USER indicate which factor factory or user is currently in use see fig 23 3 Press the meter and hold the RESET key The METER shows FIELD and the Batch Total is zero The me...

Page 6: ... basis of several performed dispensing operations this can be corrected by applying a correction to the current calibration factor as shown below New cal Factor Old Cal Factor x 100 E 100 Example Error percentage found E 0 3 value observed is 0 3 less than actual CURRENT calibration factor 1 000 New USER K FACTOR 1 000 100 0 3 100 1 000 100 0 3 100 1 003 1 Wait until the METER enters Standby displ...

Page 7: ...fixed battery symbol appears In this condition the METER continues to operate correctly but the fixed icon warns the user that it is time to change the batteries see fig 42 5 2 1 2 If meter operation continues without changing the batteries the second battery alarm level will be reached which will prevent any operation In this condition the battery icon starts to flash and is the only one to remai...

Page 8: ... 12 months from purchase date proof of which is required for any claim Weee Regulations Dispose of this product at the end of its working life in compliance with the EU Directive on Waste Electrical and Electronic Equipment WEEE When the product is no longer required it must be disposed of in an environmentally protective way Contact your local solid waste authority for recycling information Batte...

Reviews: