Manual revision 005
Section 6: Troubleshooting
SBE 63
36
Cause/Solution 3 (SeaCATs only)
: A data file with nonsense values may be
caused by incorrect instrument configuration of the SeaCAT (SBE 16
plus
V2,
16
plus
-IM V2, or 19
plus
V2):
•
Bad data may be caused by incorrect setup in the CTD. Send
GetCD
or
DS
to the CTD
to verify the CTD setup (SBE 63 is enabled for the
RS-232 auxiliary sensor channel).
•
Bad data may be caused by incorrect configuration in the instrument
.xmlcon configuration file, and/or incorrect calibration coefficients in the
instrument configuration file. Verify the settings in the configuration file.
Note:
Each SeaCAT CTD is shipped with a
configuration file that matches the
configuration of the instrument
(number and type of auxiliary
sensors, etc.) and includes the
instrument calibration coefficients.