![Sea-Bird Electronics SBE 63 User Manual Download Page 23](http://html.mh-extra.com/html/sea-bird-electronics/sbe-63/sbe-63_user-manual_1221389023.webp)
Manual revision 005
Section 4: Deploying and Operating SBE 63
SBE 63
23
Status
Command
(continued)
GetCC
Get and display calibration coefficients,
which are initially factory-set and should
agree with Calibration Certificate shipped
with SBE 63. Example below includes,
where applicable, command used to
modify parameter.
•
<CalDate> is number of days since
January 1, 2000.
•
SetRefSal=
and
SetRefP=
are set by
user; see description below in
Calibration Coefficients
section of
these command descriptions
Example:
(user input in bold)
S>
getcc
<CalibrationCoefficients DeviceType = 'SBE063' SerialNumber = '0013'>
<Calibration format = 'TEMP1' id = 'Temperature'>
<SerialNum>06300013</SerialNum>
<CalDate>04826</CalDate>
<TA0>+5.593530e-04</TA0>
[
SetTA0=
]
<TA1>+2.756939e-04</TA1>
[
SetTA1=
]
<TA2>-1.453041e-06</TA2>
[
SetTA2=
]
<TA3>+1.535286e-07</TA3>
[
SetTA3=
]
</Calibration>
<Calibration format = 'OX1' id = 'OptOxygen'>
<SerialNum>06300013</SerialNum>
<CalDate>04826/CalDate>
<A0>+8.900000e-01</A0>
[
SetA0=
]
<A1>-1.520000e-03</A1>
[
SetA1=
]
<A2>+0.00</A2>
[
SetA2=
]
<B0>-1.820000e-01</B0>
[
SetB0=
]
<B1>+1.00</B1>
[
SetB1=
]
<C0>+1.737200e-01</C0>
[
SetC0=
]
<C1>+6.545500e-03</C1>
[
SetC1=
]
<C2>+8.945700e-05</C2>
[
SetC2=
]
<E>-2.600000e-03</E>
[
SetE=
]
<SOLB0>-6.245230e-03</SOLB0>
[
SetSolB0=
]
<SOLB1>-7.376141e-03</SOLB1>
[
SetSolB1=
]
<SOLB2>-1.034100e-02</SOLB2>
[
SetSolB2=
]
<SOLB3>-8.170830e-03</SOLB3>
[
SetSolB3=
]
<SOLC0>-4.886820e-07</SOLC0>
[
SetSolC0=
]
<REFSALpsu>+0.00</REFSALpsu>
[
SetRefSal=
]
<REFPRESSdbar>+0.00</REFPRESSdbar>
[
SetRefP=
]
</Calibration>
</CalibrationCoefficients>
<Executed>