Manual revision 013
Appendix III: RS-485 Interface
SBE 38
28
ID
Commands
Only one SBE 38 can be online when sending these commands.
ID?
Display SBE 38 ID (ID = ii, where ii= 0-99) and
whether ID is required as a prefix for commands to
individual SBE 38s (see
IDReq=
).
*ID=ii
Set SBE 38 ID to ii, where ii= 0-99. *
ID=
must be
sent twice, because the SBE 38 requests
verification.
If more than one RS-485 instrument
is online when sending this command, all
instruments online will be set to same ID
.
All Other Commands
All other commands (status, setup, sampling, and coefficients) are listed
in
Section 4: Deploying and Operating RS-232 SBE 38
and in
Appendix IV: RS-232 Command Summary
.
Notes on use of these commands for an SBE 38 with RS-485 interface:
•
Effect of
IDReq=
:
If
IDReq=Y
, precede these commands with
#ii
(ii = 0 – 99) to direct a
command to a particular SBE 38.
If
#iiIDReq=N
, do not precede these commands with
#ii
. This setting
works only for a system with one SBE 38.
•
The RS-485 interface cannot accommodate multiple instruments
transmitting real-time data at the same time. Therefore, do not set
AutoRun=Y
(start sampling continuously when power is applied) for a
system with more than one RS-485 instrument on a pair of wires.
Similarly, do not attempt to send
Go
to multiple SBE 38s on a pair of
wires, as the system cannot transmit a command to one instrument while
another instrument is transmitting real-time data.
•
If planning to sample continuously:
If
AutoRun=Y
(sample
continuously when power is applied) or using the
Go
command to start
sampling continuously,
set NAvg= (A/D cycles to average per sample)
to a value greater than or equal to 30.
Data Formats
Converted Data (Format=C)
Output from continuous sampling:
ttt.ttt
Output from
DATAii
,
#iiTS
,
TS
,
#iiSH
,
SH
,
#iiSL
,
SL
,
#iiSLT
, or
SLT
:
ii, sssss, ttt.ttt
where
:
t = temperature (°C, ITS-90)
(number of digits to right of decimal point is defined by
Digits=
)
ii = ID (0 – 99)
sssss = SBE 38 serial number
Raw Data (Format=R)
nnnnnn.n
where:
n = counts
Notes:
•
Each line of output is followed by
a carriage return and line feed.
•
For converted data, leading zeros
for temperature output are
suppressed, except for one zero
to the left of the decimal point
(for example, 0.1034).
Note:
IDReq=Y
does
not
require a
#ii
prefix.
IDReq=N
does
require a
#ii
prefix
(
#iiIDReq=N
).
32