2.0 Safety and Preparation for Use 

Page 3 

The  power  return  (pin  4)  is  NOT  connected  to  the  instrument  case  ground  internally, 
however both ground connections pin 4 and pin 6 are available at the DC power connector 
and may be connected together at this point.  
 

Requirements for the external DC power supply  

The following specifications should be used to ensure optimum performance: 
DC Supply voltage    

+12 to +36 VDC, 2 Amps  

Line regulation  

 

+/- 0.05% for a 10% line change  

Load regulation  

 

+/- 0.05% for a 50% load change  

Output ripple  

 

< 5 mV peak-to-peak  

Pin configuration 

 

Same as RM12BRD-6PH(71) DC connector on the back panel  

 
Verify  that  the  connector  from  the  DC  power  supply  to  be  used  has  the  correct  pin 
configuration  mentioned  above.  Do  not  apply  AC  voltage  to  the  DC  power  connector. 
Failure to follow these directions may cause injury or death to personnel, cause irreparable 
damage to the instrument and voids all warranties.  
 
If  you  provide  DC  power  to  the  unit,  detaching  the  DC  power  cord  is  the  only  option  of 
disconnecting the unit from the DC mains supply. Make sure you have access to the rear 
panel or provide an external accessible DC disconnect means for the instrument. 
 

2.2 Instrument Safety and Preparation for Use 

 

The  DA-1G  is  designed  to  distribute  RF  signals  with  a  frequency  of  700  to  1100  MHz. 
Output levels below +7 dBm will trigger a fault condition which can be monitored from the 
front panel. Input signals must be kept below +20 dBm as greater power levels will damage 
the unit and void all warranties. 
 

Absolute Maximum Ratings 

Input RF Power  

 

 

+20 dBm Maximum 

Reverse RF Power   

 

+20 dBm Maximum 

Voltage at the RF Input  

   

+50 VDC Maximum  

Voltage at the RF Output     

+50 VDC Maximum 

Input DC Supply Voltage   

+ 36 VDC 

Storage Temperature 

 

-10 to +75 ºC 

Operation Environment  

 

0 to +50 ºC 

Summary of Contents for DA-1G

Page 1: ...DA 1G 1 GHz DISTRIBUTION AMPLIFIER OPERATING MANUAL SPECTRADYNAMICS INC 1849 Cherry St Unit 2 Louisville CO 80027 Phone 303 665 1852 Fax 303 604 6088 www spectradynamics com SPECTRADYNAMICS INC...

Page 2: ...DA 1G 1 GHz Distribution Amplifier Operating Manual Copyright 2022 SpectraDynamics Inc All rights reserved DA 1G 2022MD...

Page 3: ...Safety and preparation for use 2 2 1 Electrical 2 2 2 Instrument 3 3 0 Front panel description 4 4 0 Back panel description 5 5 0 Installation 6 6 0 Operation 7 7 0 Troubleshooting 8 8 0 Specification...

Page 4: ......

Page 5: ...level monitor that compares the twelve output power levels to a preset threshold of 7 dBm If the signal on any output drops below this threshold the monitor LED located on the front panel will turn of...

Page 6: ...Only replace fuse with the same type and specifications AC Power The instrument has a detachable three wire power cord for connection to a grounded AC power source The enclosure of the unit is direct...

Page 7: ...low these directions may cause injury or death to personnel cause irreparable damage to the instrument and voids all warranties If you provide DC power to the unit detaching the DC power cord is the o...

Page 8: ...is installed the DC Power LED will be on when DC power is applied to unit Monitor LED The monitor LED will be on if all RF outputs are greater than 7 dBm RF signal levels less than 7 dBm will trigger...

Page 9: ...1G is set up to operate with both AC and DC power sources at the same time the DC power is used as backup power in case of AC power outages INPUT A RF Signal within the range of 700 MHz to 1100 MHz ma...

Page 10: ...the AC POWER entry module on the rear of the enclosure and connect the power cord The amplifier units that are acquired with the DC operation option are shipped with a DC connector part number RM12BPE...

Page 11: ...nuous operation Once AC power is supplied to the DA 1G the LED on the front panel labeled AC POWER will turn on If you also applied DC voltage the LED labeled DC POWER will also turn on The DA 1G is d...

Page 12: ...and DC power cords Remove the top cover Check the DC power fuse and power cord If the fuse is blown replace with same type and rating Please contact SpectraDynamics if any of the fuses blows again or...

Page 13: ...ironment 0 to 50 C Humidity 5 to 95 Non condensing PARAMETER CONDITIONS MIN TYP MAX UNITS Output power level 1 dB compression 13 dBm Bandwidth 1 dB 800 1100 MHz Gain 1 GHz 0 5 0 5 1 5 dB Impedance Inp...

Page 14: ...ion or any liability for direct indirect or consequential damages or delay Any improper use operation beyond capacity substitution of parts not approved by SDI or any alteration or repair by others in...

Page 15: ...ertain voltage limits 2014 30 EU The Electromagnetic Compatibility Directive and its amending directives 2011 65 EU of the European Parliament and of the Council of 8 June 2011 on the restriction of t...

Reviews: