SDI DA-1G Operating Manual Download Page 12

7.0 Troubleshooting 

Page 8 

Do not attempt to service or adjust the instrument unless another person, capable of 
providing first aid or resuscitation, is present.  
 

If there  are problems that  cannot be resolved by the troubleshooting steps below please 
contact technical support.   
 
Technical Support   
Tel: +1 (303) 665-1852 , Fax: +1 (303) 604-6088  
[email protected], www.spectradynamics.com  
 

AC Power LED does not turn on. 

Disconnect AC and DC power cords. Remove the top cover. Check the AC power fuse and 
power cord. If the fuse is blown replace with same type and rating. 

 

 

DC Power LED does not turn on. 

Disconnect  AC  and  DC  power  cords.  Remove  the  top  cover.  Check  the  DC  power  fuse 
and power cord. If the fuse is blown replace with same type and rating.  
 
Please contact SpectraDynamics if any of the fuses blows again or if the event that caused 
the fuse to blow is not known. 
 

Monitor LED is off.  

Check to see if the RF signal provided to the instrument is greater than +7 dBm.  
 
Check to see if a signal is present at all outputs of the signal distribution module.  
 
The Monitor LED will remain off with a power lever under +7 dBm, but will still provide five 
copies of the RF input signal. If the instrument is providing the copies of the input signal 
you may continue using the DA-1G.  
 
If the power level of the RF signal provided is greater than +7 dBm and the Monitor LED 
remains off, the instrument will have to be returned for repair.  

Summary of Contents for DA-1G

Page 1: ...DA 1G 1 GHz DISTRIBUTION AMPLIFIER OPERATING MANUAL SPECTRADYNAMICS INC 1849 Cherry St Unit 2 Louisville CO 80027 Phone 303 665 1852 Fax 303 604 6088 www spectradynamics com SPECTRADYNAMICS INC...

Page 2: ...DA 1G 1 GHz Distribution Amplifier Operating Manual Copyright 2022 SpectraDynamics Inc All rights reserved DA 1G 2022MD...

Page 3: ...Safety and preparation for use 2 2 1 Electrical 2 2 2 Instrument 3 3 0 Front panel description 4 4 0 Back panel description 5 5 0 Installation 6 6 0 Operation 7 7 0 Troubleshooting 8 8 0 Specification...

Page 4: ......

Page 5: ...level monitor that compares the twelve output power levels to a preset threshold of 7 dBm If the signal on any output drops below this threshold the monitor LED located on the front panel will turn of...

Page 6: ...Only replace fuse with the same type and specifications AC Power The instrument has a detachable three wire power cord for connection to a grounded AC power source The enclosure of the unit is direct...

Page 7: ...low these directions may cause injury or death to personnel cause irreparable damage to the instrument and voids all warranties If you provide DC power to the unit detaching the DC power cord is the o...

Page 8: ...is installed the DC Power LED will be on when DC power is applied to unit Monitor LED The monitor LED will be on if all RF outputs are greater than 7 dBm RF signal levels less than 7 dBm will trigger...

Page 9: ...1G is set up to operate with both AC and DC power sources at the same time the DC power is used as backup power in case of AC power outages INPUT A RF Signal within the range of 700 MHz to 1100 MHz ma...

Page 10: ...the AC POWER entry module on the rear of the enclosure and connect the power cord The amplifier units that are acquired with the DC operation option are shipped with a DC connector part number RM12BPE...

Page 11: ...nuous operation Once AC power is supplied to the DA 1G the LED on the front panel labeled AC POWER will turn on If you also applied DC voltage the LED labeled DC POWER will also turn on The DA 1G is d...

Page 12: ...and DC power cords Remove the top cover Check the DC power fuse and power cord If the fuse is blown replace with same type and rating Please contact SpectraDynamics if any of the fuses blows again or...

Page 13: ...ironment 0 to 50 C Humidity 5 to 95 Non condensing PARAMETER CONDITIONS MIN TYP MAX UNITS Output power level 1 dB compression 13 dBm Bandwidth 1 dB 800 1100 MHz Gain 1 GHz 0 5 0 5 1 5 dB Impedance Inp...

Page 14: ...ion or any liability for direct indirect or consequential damages or delay Any improper use operation beyond capacity substitution of parts not approved by SDI or any alteration or repair by others in...

Page 15: ...ertain voltage limits 2014 30 EU The Electromagnetic Compatibility Directive and its amending directives 2011 65 EU of the European Parliament and of the Council of 8 June 2011 on the restriction of t...

Reviews: